EUROPEAN SOUTHERN OBSERVATORY
La Silla - Science Operation Department
|
HARPS major interventions
|
| Version: 1.1 2007-05-08 Gaspare Lo Curto
|
Interventions
Althougth the instrument is very stable and requires minimum maintenance,
at times major interventions are requested.
This page is intended to log them.
14-01-2005 -> 22-01-2005:
- vacuum vessel and CFC pressurized;
- CFC opened;
- PKR251 high sensitivity pressure gauge changed;
- high sensitivity HPT100 pressure gauge inserted in the line for backup;
- set of valves installed to avoid pressurizing the vacuum vessel next time
an intervention on the vacuum gauges is needed.
- Broken wires to some PT100 sensors repaired.
Collateral effects:
- "Warm" pixel at (x,y)=(1700,1622) became "hot" giving 10000 counts
at bias and affecting the column above, making it useless. This column
crosses order # 122 on fiber B (reference) and is included in the "bad" column li
st of the pipeline;
- the vacuum vessel cannot reach pressure lower than 10E-3 mbar even
after 5 hours pumping.
10-09-2005 -> 12-09-2005:
- temperature sensor 43 (isolation box secondary) changed of location
from the air flow tube to below the vacuum vessel;
- reference sensor changed from 42 to 43;
- temperature of the reference sensor set to 16.2.
09-10-2006 -> 10-10-2006: