Procedure: SOFI Mask/Slit Wheel - Align
Author: Leo Vanzi & V. Doublier
Release: 1.0/ June 22, 2002

The Slit/Mask wheel is ONLINE but mis-aligned

A- Re-initialise the wheel

- Open the Motor Device control Panel (click on title bar of the function)

- Select HW init: if the wheelk fails to initialise see: procedure for Slit/Mask wheel initialisation failure , if the wheel initialise then:

B - Move the wheel to Slit_0.6:

- On the OS: click MORE... button and enable "draw RTD Object"

- If there is less than 1pixel difference over the entire length of the detector: resume normal operations

- if the position is wrong more than 1pixel, try to go to another position and come back to original position

- If the mis-alignment increase when you come back to the 0.6 arcsec slit, call for help

- if the mis-alignment is the same, you will have to re-align the wheel

C - Aligning the Slit/Mask wheel

- Run the template SOFI_spec_tecAlignSlit, which can be found in OBD/CALIB in BOB.

- if the misalignment is less than 1 pixel: do not change the encoder position and resume normal operations, if not chnage encoder position and ...

- Repeat until mis-alignment is better than 1pixel, then perform a Fall Back initialisation

- Resume normal operations, and REPORT Both failure conditions and wsofi LogMon error messages in Remedy

 

 