European
Southern Observatory
New Technology Telescope |
| General : | |||||
| Observations : | |||||
| Operations : |
![]()
| Procedure: | SOFI Mask/Slit Wheel - Align |
| Author: | Leo Vanzi & V. Doublier |
| Release: | 1.0/ June 22, 2002 |
The Slit/Mask wheel is ONLINE but mis-aligned
A- Re-initialise the wheel
- Open the Motor Device control Panel (click on title bar of the function)
- Select HW init: if the wheelk fails to initialise see: procedure for Slit/Mask wheel initialisation failure , if the wheel initialise then:
B - Move the wheel to Slit_0.6:
- On the OS: click MORE... button and enable "draw RTD Object"
- If there is less than 1pixel difference over the entire length of the detector: resume normal operations
- if the position is wrong more than 1pixel, try to go to another position and come back to original position
- If the mis-alignment increase when you come back to the 0.6 arcsec slit, call for help
- if the mis-alignment is the same, you will have to re-align the wheel
C - Aligning the Slit/Mask wheel
- Run the template SOFI_spec_tecAlignSlit, which can be found in OBD/CALIB in BOB.
- if the misalignment is less than 1 pixel: do not change the encoder position and resume normal operations, if not chnage encoder position and ...
- Repeat until mis-alignment is better than 1pixel, then perform a Fall Back initialisation
- Resume normal operations, and REPORT Both failure conditions and wsofi LogMon error messages in Remedy
| Send
comments to: NTT Webmasters <nttweb@www.ls.eso.org> Last Update: MMM DD, YYYY (WHO) |