| General : |
|
|
|
|
|
| Observations : |
|
|
|
|
|
| Operations : |
|
|
|
|
|

MOS
- if emmi punch still doesn't work punching has to be performed @
3.6
What follows is not any more available.
- take emmi MOS plates in EMMI closet.
- goto 3.6
- login as mos@w3p6ins in 3.6 mos room
- start mos tool panel mosuifMOS
- online mos
- click on Move STAP after defining position #n just above this key.
- load a blank plate into position #n
- DO NOT USE mos tool panel to punch EMMI plates as it only allows
EFOSC range
but open a xterm window
- cd tmp
- ftp in *.mask
files from where they have been created with xm.
- type
punch #n <file>.mask
- if you have several plates to punch
- you can create a script looking like
punch 2 firtstfield00.mask
punch 3 firstfield01.mask
punch 4 secondfield03.mask
punch 5 lastfield00.mask
and call it for instance allmasks
- then , after loading blank plates in positions 2,3,4,5
chmod +x allmasks
allmasks
which will punch all masks in row whitout need to stay there
to change the plates
- if it doesn't work , reload punch
and Make3.6 scriptps into
tmp/ after a cmmCopy xmospan
(src/)
- click on Move STAP after defining position #n just above this key.
- unload punched plate from postion #n
- bring all punched plates to EMMI room.
- Press EMMI emergency stop button at the center of EMMI cover
- start EMMI Config panel from NTT menu> INS
- load punched masks sequentially
- move to load starplate
- load starplate
- declare MOS#<n>
- enter in position
- Skip this step if you come with already punched plates.
- Get the requested punching head from cupboard #1.
- Open the red EMMI cover.
- Unplug the connecting cable of the punching head (function #25).
- Rotate the button on the punching head by 90 deg to release the
punching head, then pull it out carefully.
- Carefully put in the new head and once in position rotate the button
on the punching head by 90 deg to lock it in position.
- Plug in the connecting cable of the punching head (function #25)
- Install MOS plates into Starplate wheel like normal
slits (Except Declaration and Alignment).
- start mos tool panel from os EMMI panel
- online mos , load mask and start.
- Close the EMMI cover.
- Declare MOS plates into EMMI configuration panel.
- Adjust punching offset (MOS operation Manual
, 2.1.3)
- Click Backup under File submenu.
- Release the emergency stop button.
![[Top of Page]](../../buttons/up_arrow.gif)