Starplate Setup
The starplates are mounted on a wheel offering 5 positions, one being always
FREE. This unit corresponds to the EMMI function STAP. The starplate can be
of different types: slit, MOS plate, pinhole, mask, .....
- Press EMMI emergency stop button at the center of EMMI cover.
- Get the box containing all the possible starplates from cupboard #1 or from
the platform table.
- Remove the large red cover from the front of EMMI.
- From the EMMI configuration panel, select the desired position and click
move to load button of STAP.
- Loosen the four clamping screws and remove the plate.
- The new starplate has to be inserted with the cut corner inside upwards.
- Push the plate fully into position to have it well positioned on
the reference pin.
- Tighten the four screws with a moderate torque.
- Put back the red EMMI cover.
- Release the emergency stop button.
- Declare installed slits into EMMI config panel.
- Run RILD_img_tec_SlitAlign to align installed slits
- Check that slit position and slit encoder offset are
modified consistently on EMMI config panel.
- Click Backup under File submenu.