The daily `health checks' on SOFI require the operator to revise the Status Panels of the instrument, in order to determine whether the instrument is in its normal operational state or if there are any cryogenic or vacuum problems imminent.

To start the Status Panel on `wsofi', proceed as follows:

Table 1 indicates the allowed range of the values given in the panels:
 
 
Sensor
Panel
Normal Range when COLD
Unit
Cool-Down
Temperature
75.0 - 85.0
K
Camera Unit
Temperature
65.0 - 80.0
K
Radiation Shield
Temperature
90.0 - 115.0
K
Mirror
Temperature
65.0 - 80.0
K
Collimator
Temperature
65.0 - 80.0
K
Warm-Up
Temperature
65.0 - 80.0
K
Sorption Trap
Temperature
130.0 - 160.0
K
Aperture Wheel
Temperature
70.0 - 80.0
K
Sensor #1
Temperature
65.0  +/- 0.1
K
Vacuum Pressure
Pressure
< 10 e-6
mbar
Signals
Signals
All Inactive -> Blue
 logic
 
Table 1

 Any deviation from the above should be reported immediately to the IRC.