Focus Wedge Aligment
Focus wedge is supposed to be mounted permanently
in position 8 and should never be removed.
If however you have to mount it
, proceed as follows : anyhow its alignment MUST be checked at every
setup :
- Press EMMI emergency stop button at the center of EMMI cover.
-
Remove the black side cover of EMMI red very carefullly to avoid
to damage the grism motor. - From the EMMI configuration panel, select grim
pos 8 and click "move to load".
- Install focus wedge in position.
-
From EMMI configuration panel, select grism pos 8, select grism name
Focus wedge
and click enter grism. - Click Backup
under File submenu.
- Check that pinhole is mounted in Starplate
wheel.
- Release the emergency stop button.
- To align focus wedge,
load and run RILD_img_tec_FocWdgAlign.obd.
- First make sure instrument is in focus running RILD_img_tec_SetFoc.
- Remind
that RILD_img_tec_FocWdgAlign will ask you to click left image in PCO MIDAS window.
- Check that EMMI configuration panel is modified accordingly.
- Click
Backup under File submenu.