Measuring the signal current (voltage) of a IK320 measuring system
The signal amplitude at the pins 2 and 3 is determined by the current signal. The current/voltage conversion takes place with the factor 112mV/uAss.
|
Pin |
signal assignation |
|
1 |
U0-GND |
|
2 |
U90 |
|
3 |
UOO |
|
4 |
RI |
|
TRIG.RI |
|
|
6 |
GND |
Example: UOO = 1.68 Vss (measured on pin 3 of P1 or P2)
I00 - U00/Factor = 1,68Vss/112mV/uA 15 uA
2. Measurement over the AD converter of the IK320
To access the AD converter see description chapter 5.2 data.
The analog values are stored under the Base address + $F8... $FE.
With slow speed now the MIN- and MAX- values of the signal amplitude are determined.
The current signal is calculated as follows: The amplitude Ass is the total amount of the MIN and of the MAX value. An AD transducer increment has 0,6lmV, from which the amplitude in Vss can be calculated. The amplification factor of the input stage altogether is 540mV/uA.
Example:
MAX AD-Converter: $67C8 arithmetic shift right 2 = $19F2
MIN AD— Converter: $984C arithmetic shift right 2 = $E613
Total amounts: $33df
Decimal: 13279
Uss = 0,61mV * 13279 = 8,1 Vss
Iss = 8,1 Vss / 540 mVuA = 15 uAss