To replace the head, protect first the head and the strip by introducing a plastic foil between the head and the strip. Unscrew the bolts holding the head in place and withdraw it. Install the new head in its support, aligning it so its base is parallel to the strip. Adjust the distance using a plastic gauge of 0.7mm. Connect the PWM8 encoder analyzer (Heidenhain) to a 12 to 30 volts power (special connector). Connect the head to the INPUT connector of the PWM8 encoder analyzer by the special adapter cable that matches the head and the PWM8 box.
Note: There are two plug in modules for the PWM8 box: one for 11 uA type head and one for voltage head. The 3.6m heads are "voltage" type.Switch on the PWM8. Confirm that the option to supply +5 volts to the encoder head is on (Um:on) Rotate the telescope axis and watch the reading amplitude of both signals (0° and 90°). The maximum amplitude is obtained by the optimal alignment of the head respect to the strip marks (fig. 1). To adjustment, release a bit the bolts that hold the head (A) and rotate it slightly searching the maximum signal amplitudes. Match the amplitude of both signals by trimming the “Compensate” potentiometer in the head electronics and then adjust the amplitude to 1 Vpp with the “Strom” potentiometer.
TVl, TV2 and Phase Adjustment:
- Set TV1 and TV2 potentiometers in the encoder PC board, to their central position. While moving the telescope in slew speed, tip and tilt the head holder with the adjusting bolts trying to obtain a phase reading below 20° (fig. 4)
- Adjust potentiometers in the encoder head to obtain TV1 and TV2 below 20° (fig. 4)
- Confirm that the phase between 0° and 90° signals. The PWM8 box displays the departure from the 90° degrees phase in degrees.
- To align the phase, change the head-strip tilt-angle on the head support adjusting screws. Match the amplitude of both signals by trimming the “Amplitude” potentiometer in the head electronics (Fig. 2) and then adjust the amplitude to 1 Vpp with the “Strom” potentiometer.
- As a final check, repeat the whole procedure to verify that all the parameters are OK
Tight the bolts and watch that the signals stay at the maximum.
Fig. 1 Fig 2.
Fig. 3
| Reference mark adjustment (Ri signal) | |
![]() |
Use an oscilloscope with
three channels and memory. Connect the 0° and 90° signals
from PWM8 box to channels A and B respectively
and add them. The third input is the reference
itself. If you do not have a three-channel oscilloscope, you may add the
0° and 90° signals connecting them via two 200K ohm
resistors to one channel. Connect the other scope channel to the reference
mark signal (zero pulse). Set the trigger source to the reference mark signal,
trigger mode to single and adjust the trigger level to +0.4 VDC, negative
slope.
Move the telescope and wait for the arrival of a zero pulse. The phase difference of the zero pulse
(negative peak) respect to the sum of 0° + 90° signals
should be less than 20° (fig.
4). To adjust the zero pulse phase, tilt the encoder head using the
adjusting bolts in the head support. Don't get confused with the "spring
tension" bolts! |
|
Fig. 4
|
Repeat all the previous adjustment |