TCS and Autoguider: Basic Procedures
Presetting the Telescope
Entering object coordinates
It is possible to enter
object coordinates in two ways: directly in the TCS panel, or through a
catalog.
To enter coordinates
directly on the TCS panel, click on the input line located on the Presetting
portion of the TCS control panel, delete any previous coordinates and enter
the coordinates of the object, RA first and declination last, leaving a
space between RA and DEC and no spaces within each one (HHMMSS -DDMMSS).
Click on the input line on the right, labeled Epoch and enter the coordinate's
epoch.
If you prefer you can
make a catalog with different objects so you don't have to enter the coordinates
every time you change objects.
To do so, open a text
editor in the TCS workstation and write the coordinates of the objects,
one per line with the following format:
Name HHMMSS.ss -DDMMSS.ss
Epoch Prop.Mo.RA Prop.Mo.DEC
The sign in front of
the declination is optional if it is positive. The decimal numbers are
also optional, including that of the epoch. The epoch is mandatory, otherwise
it will default to 00.00. The proper motion values will default to 0.0
also if you don't enter them.
Once you have entered
the desired objects in the text file, save it as /home/tcs/catal/name.cat
replacing name with a suitable name.
Go to the TCS Control
panel and press Cat.Select on the Catalogue Handling portion
of the panel. It will show a window with the different catalogs stored
on the directory. Click on your catalog and click on OK. Now the
catalog should be loaded in the Catalog Handling portion of the
panel. Use the UP and Dwn buttons to highlight the desired
object. Once chosen, the object coordinates should be entered automatically
in the input fields above.
Presetting
Once the coordinates of
an object has been entered in the coordinates input fields, press the
Preset
button. The
Telescope Status field should show
PrstToObj.
Wait until the word
Guide show up on the same field; it indicates
that the telescope has arrived at the desired coordinates. If the telescope
doesn't move after you have pressed
Preset then it is possible that
your object is out of the telescope's safety zone; check the
Messages
window for an error message.
Acquiring a Guidestar
Once you have positioned
the telescope on the desired coordinates you will want to pick a guidestar.
To do so follow this procedure.
-
Press the Retrieve Field
on the AG Field Acquisition portion of the TCS Control Panel.
-
Wait until the acquisition
image has loaded on the Real Time Display, on the right hand monitor of
the TCS workstation and the word
Complete appears on the CCD
Stat field, on the Autoguider portion of the TCS Control Panel.
It is very important to wait for the image to load completely before
doing anything else, otherwise you risk crashing Fiera, whose restart procedure
could take twenty minutes or more.
-
If the acquisition image
doesn't begin to load on the RTD after a reasonable amount of time (1 minute
aprox), go to the AG Process workspace, kill every window there,
open an xterminal and type the command /home/tcs/m2p2/bin/agWSStart.sh.
Wait for the four text windows to appear and repeat the process. If the
four text windows do not appear it is very possible that Fiera has crashed.
Try to take a science test image; if it doesn't work go to the procedure
to restart Fiera in the Notes for Observers document located in
the control room.
-
Once the acquisition image
is completely loaded the RTD, click on the Pick Object button on
the Pick Reference Star panel on the right hand monitor. If the
panel isn't there, open the TCS menu on the RTD panel, click on
Pick
Reference Star and wait for the panel to appear. Click on a suitable
guidestar. A box should appear centered on the star. Click on the Star
To Box button on the
Autoguider portion of the TCS Control Panel.
After a little while a small window should appear on the RTD with the selected
star centered on it. The images of the guidestar should be refreshed continuesly
and the AG Stat field on the
Autoguider portion of the TCS
Control Panel should say
ACTIVE. Also the Telescope Status
field on the TCS Control Panel should say Offset and
Ofsinpr
alternatively as the autoguider starts to correct the position of the star.
If all these conditions are met, the autoguider is active and you can start
your science exposure. Otherwise, turn off the autoguider clicking the
Off button and try to start it again with the Star to Box
button. If that doesn't work, try pressing the OFFSET button, on
the Virtual Handset on the TCS Control Panel before starting the
autoguider again.
-
Sometimes it happens that
images of the guidestar stops arriving to the autoguider. If you see the
autoguider window frozen that is the case. Go to the AG Processes
and repeat the process of killing and restarting the text windows there,
as described above. If that doesn't work, Fiera could have crashed. Test
with a short science exposure and restart Fiera if necessary.
-
After Restarting the text
windows on the AG Processes workspace, or after restarting Fiera,
the autoguider will lose the position of the guidestar and the acquisition
process should be started from the begining.
Performing a Combined
Offset
A combined offset is the
action of moving the telescope
and the autoguider window a small
amount, so the autoguider can keep track of the guidestar and the time
consuming procedure of acquiring a guidestar can be avoided for small offsets.
To perform a combined
offset follow this procedure:
-
Turn off the autoguider,
clicking on the button Off on the Autoguider portion of the
TCS Control Panel.
-
Check (if it is not already
checked) the button Combined Offset on the Virtual Handset
portion of the TCS Control Panel.
-
Enter the amounts of the
desired offsets in RA and Dec on the input fields below the Virtual
Handset and press the Apply button to load the values into the
software. This input fields only accepts positive numbers; the direction
of the offsets is given by the button of the handset you press.
-
Press the apropiate buttons
on the virtual handset to perform the offset and wait for the word Offset
on the
Telescope Status field.
-
Press the Star to Box
button on the
Autoguider to switch on the autoguider again.
IMPORTANT: Always turn
off the autoguider before performing a combined offset; otherwise you risk
crashing Fiera and the procedure to restart it can take more than twenty
minutes.
Changing filter
The autoguider shares
the same filter frame with the science mosaic, so, when you change filter
the guidestar won't be exactly in the same place than with the previous
filter due the differences in refraction coefficient of the different filters.
You must turn off the autoguider before changing filters, wait for the
new filter to be in place, and restart the autoguider with the button
Box
to Star. This way the telescope won't jump to recenter the guidestar
when you start the autoguider.