2p2-m Telescope
|
|||||
|
The plan we ar using has two
components: Instrument Calibration and Standard Calibration.
The notes on this web page are based on (but supercede) the short document
WFI Calibration Plan
of February 2001.
The public WFI Calibration
Plan web page provides links to the results.
Instrument Calibration aims solely to provide data to
quantify instrument behaviour for the information of the Team.
It does not aim to supply calibrations for use in reduction.
However, these Instrument Calibration
data are available to users through the links below.
The analysis of Instrument Calibration data is
undertaken mostly by us.
Instrument Calibrations are grouped in terms of Short, Medium and Long-Term
Calibrations:
Instrument Calibration:
| Short-Term Calibrations | Status | Frequency | Contact | Execution Notes | Analysis Notes |
|---|---|---|---|---|---|
| CCD Bias Levels, Gains, Read-Noises (Daily Health Check) | Running | Daily | F. Selman | * How to run it * Daily Health Check OB | * Daily Health Check analysis script * Daily Health Check plotting script |
| Image Quality Database | Running | Hourly (at night) | L. Germany | *
Image quality input page | * README notes * Script to read input from web page * Script to plot results requested via web plot interface |
| Pointing Reproducibility (Nightly Health Check) | Scripts Ready | Nightly | L. Germany | * How to run it * OB for Template Field No. 1 | * README notes * nightlyhealthcheck.pl script to process data |
| Image Quality Across Mosaic (Nightly Health Check) | Scripts Ready | Nightly | L. Germany | * How to run it * OB for Template Field No. 1 | * README notes * nightlyhealthcheck.pl script to process data |
| Differential Throughput Measurements (Nightly Health Check) | Scripts Ready | Nightly | L. Germany | * How to run it * OB for Template Field No. 1 | * README notes * nightlyhealthcheck.pl script to process data |
| (Weekly Health Check) | Current data not transferred | Weekly | F. Selman | * How to run it * Weekly Health Check OB | . |
Standard Calibration frames are science calibrations taken
on a routine basis in order to determine photometric zeropoints in
UBVRI.
They consist of three or four Landolt standard star fields taken throughout
the night in all or part of UBVRI, with associated dome, twilight flats
and biases.
They are taken on each photometric night in which Service Mode
programmes of matching UBVRI are also being observed.
The analysis of Standard Calibration data is not undertaken by
MET but by the users who need it to calibrate their Service Mode data.
Standard Calibration data are distributed to the relevant users as
calibration frames with their programme data. Service Mode users requiring
UBVRI photometry need not request them and are not charged the time for them.
They also contribute (in part) to the Instrument Calibration.
Any science calibrations outside of our Standard Calibration plan
(e.g. flux calibration in H-alpha; more comprehensive UBVRI
photometry; narrowband twi-light flats)
are termed Non-standard Calibrations, and must be
requested through Observation Blocks. Users requiring such calibrations
are charged the time taken by them. These calibrations are not
a part of our Calibration Plan.
| Standard Calibrations | Status | Frequency | Execution Notes |
|---|---|---|---|
| Landolt Fields | Running on SM nights | 3 hourly | UBVRI or some subset as the programmes require |
| Twilight Flats | Running on SM nights | Daily | UBVRI or some subset as the programmes require |
| Dome Flats | Running on SM days | Daily | UBVRI or some subset as the programmes require |
| Bias Frames | Running on SM days | Daily | UBVRI or some subset as the programmes require |
|
|
|||