SUSI2 Problems

Version: 1.0 2007-01-18 M. Stefanon

Image quality

SuSI2 IQ has shown to be quite sensitive to external factors, especially when working at the maximum resolution. Below is a typical image of elongated star.

Elongated star

Two main causes could be identified so far: A more detailed description of these two factors is explained hereafter.

Focus Offset

This issue actually comprises three independent factors:

Filter Offset

During some technical nights in 2005-2006 an attempt to derive the offset for the standard filters with respect to the R#813 was done. The main difficulty in the execution of the tests was the small range in temperatures available. Such tests should be probably done in winter, preferably close to twilights, when the maximum gradient of temperature is present. Moreover after the first session of measurement, it was evident that temperature variations were playing the dominant role in the determination of the offsets.

The measurements consisted in R#813 and B#811 focus sequences (RBRBRB..) for a total amount of 20 times. At the beginning and end of each sequence, the temperature of M2 was also noted.
The plot below shows the dependence of the focus position on the temperature variation between two consecutive focus sequences. Namely, Tau is the difference of M2 consecutive temperatures, while Delta offset is the difference between the current filter offset and the previous one, as measured by midas. In this way, the constant term in the linear fitting gives the double of the filter offset value.

Filter offset

The computed value of 0.0014 for the filter focus offset between R and B means it can be considered negligible. (RMS???)

A confirmation of the negligible filter offsets arrived via private communication from O. Hainaut, in which he informed us that SuSI2 filters have been designed in order to have the same optical depth, meaning no filter offset should be necessary when passing from one filter to another.

Guide Probe Focus Offset

During normal operations, it is common that, even after doing Image Analysis, image quality is still not as good as one would expect.
This happens because the guide probe of side A is not exactly at the same distance as SuSI2 detector, resulting in that the best focus position measured by IA will not be the one required by SuSI2 detector. For SuSI2, the dependence of this focus offset on the defocus value given by IA has already been measured (although not recently). The formula, which can also be found in the SuSI2 Survival Guide, is:

Foc_susi = Foc_IA - def / 1030

where Foc_susi is the desired M2 value giving the correct focus for SuSI2, Foc_IA is the M2 position computed by IA, while def is the defocus value in the IA panel.

Temperature dependence

The filter offset plot shows also a temperature dependence on filter offset. However, this has never been measured accurately. TIOs are usually correcting for temperature variation using an empirical (??) formula:

Foc_new = Foc_old - 0.08 * Delta_T,

with Delta_T = T_old - T_new, T in Celsius degrees and focus positions are im mm (as in the TCS panel). While for EMMI and SofI small (~0.1 C) temperature variations can be tolerated, SuSI2 is extremely sensitive to such variations, and it is strongly recommended to do the proper corrections before the beginning of each exposure. The manual procedure of focus adjustment should possibly be automatized. In this case, it is vital to keep in mind that M2 can NOT be moved during an exposure, as an adjustment of M2 position causes the star to move on the CCD, giving an elongated or wider PSF on the final frame.

Given the high sensitivity of SuSI2 to small temperature changes, and given also that the smaller is the temperature change, the most probable to happen it is, we have that, even applying all the above mentioned corrections, image quality on the final image can still be not optimum, as temperature variations during the exposure may have changed the best focus position. However, no dedicated tests have been done so far to support this idea.

As final, although quite obvious remark, all these effect can be easily taken into account by doing a through focus sequence.

Zenith Distance

Observations done close to the zenith can generate elongated stars on the frame. The maximum speed for NTT Azimuth, Elevation and Derotator motors define a forbidden zone around the zenith of about 5 degrees. However, tests have shown that when SuSI2 is used with binning 1x1, it is advisable to enlarge such a zone to at least 10 degrees (corresponding to an airmass of 1.015).

Shift of stellar positions - Astrometry

During a high accuracy astrometric program, the PI noted a difference in stellar positions between two consecutive epochs which could not be explained by proper motions of the stars in that field. Below the three plots originally sent by the PI, clearly showing the difference in positions between observations done in February and April. All the plots refer to CCD#46.

Comparison January/Febraury Comparison January/April Comparison April/May

The first and long lasted hypothesis was that SuSI2 image quality was the main cause for this. However, direct measurement of IQ showed no direct correlation with the problem. In the plots below IQ of the stars in the filed is represented. The same images obtained by the PI were used.

IQ January IQ February IQ April IQ May

New data was taken the year after and, according to PI email exchange, the stellar positions are in agreement with the previous January and February runs. The cause of the problem was identified in the different method used for IA: while for January and February runs, IA was done in centerfield, in April and May it was done at the border (as it is usually done for the other NTT instruments). But after those runs, IA was switched back to centerfield.