Measure of the Convertion
Factor |
Author :
|
SOFI team |
Release :
|
12/09/99 : update
06/01/99 : creation |
The convertion factor (e/ADU) of SOFI is measured every month. This is done
with the template SOFI_det_tec_Gain.obd .
The results are stored in the file /vltdata/SPC/Sofi_det_tec_Gain.SPC
on wsofi, the last MIDAS images are saved in /home2/sofi/midwork
always on wsofi.
Make sure MIDAS for PCO is running on wsofi.
Make sure SUSI M4 mirror is in position SOFI.
Load the template SOFI_det_tec_Gain.obd
from /data/SOFI/SYSTEM/COMMON/OBD/CALIB
into Bob and run it. The calibration unit will be automatically moved in. You
will be asked to switch the calibration lamp on, set the light intensity to
have an average of about 1000 ADU in the illuminated part of the array when
the DIT is 1.18 (default value for the first image). To get consistent results
it is CRUCIAL to use
the HALOGEN LAMP in
the CALIBRATION UNIT.
The lamp will be automatically switched off at the end of the template and
the calibration unit put out.
The value of the convertion factor in e/ADU will be displayed in the bottom
part of the BOB window at the end of the template. It takes about 30 minutes
to run the template with the default parameters.
A plot of the square of the variance versus the average ADUs will be displayed
in a MIDAS graphic window, check that no point deviates significantly from a
linear relation. Signal the presence of deviant points.