General : |
|
|
|
|
|
Observations : |
|
|
|
|
|
Operations : |
|
|
|
|
|

Procedure :
|
Collimator Focus |
Author :
|
SOFI team |
Release :
|
12/09/99 : update
06/01/99 : creation |
The collimator focus should be determine once every three months and it should
also be checked after the instrument has been thermally cycled. The focus has
to be computed for all of the SOFI filters and, besides a filter should be given,
for each of the 6 imaging modes. These long templates can be executed, for instance,
during cloudy nights or EMMI runs. The collimator focus has proven to be very
stable, even after thermal cycling. The stepper motor values for the collimator
are recoreded in the file /data/SOFI/SYSTEM/MISC/ICS/isiSystemSOFI.cfg.
Measuring the collimator focus.
Make sure MIDAS for PCO is running.
6 Observation Blocks are available in the directory /data/SOFI/SYSTEM/COMMON/OBD/CALIB
:
- SOFI_ima_tec_InstrunentFocus_LF.obd
for the "Large Field Imaging" mode ;
- SOFI_ima_tec_InstrunentFocus_SF.obd
for the "Small Field Imaging" mode ;
- SOFI_ima_tec_InstrunentFocus_SO.obd
for the "Spectroscopic Objective Imaging" mode ;
- SOFI_ima_tec_InstrunentFocus_HR.obd
for the "High Resolution Imaging" mode ;
- SOFI_ima_tec_Ins...Foc_LF_FE.obd
for the "Large Field and Focal Elongator Imaging" mode ;
- SOFI_ima_tec_Ins...Foc_SO_FE.obd
for the "Spectroscopic Objective and Focal Elongator Imaging" mode.
Each of these Observation Blocks groups a set of 21 templates to determine
the focus position for the 21 filters of SOFI.
The results are recorded in the file /vltdata/SPC/SOFI_tec_cal_InstrumentFocus.SPC.
If a focus position has changed by less than 2000 steps, do nothing. If it
is more, you'll need to edit the configuration file.
Editing the Configuration File
In the directory /data/SOFI/SYSTEM/MISC/ICS/
make a backup copy of the configuration file isiSystemSOFI.cfg.
Bring up the ICS control panel from the tools menu of the OS panel.
From the commands menu of the ICS panel set all devices OFF. The name tags
of instrument functions will turn red.
Bring up the configuration tool from the tools menu of the ICS panel.
Select the file isiSystemSOFI.cfg from
the configuration tool and make the appropriate changes.
Press the load button and wait about 20 seconds for a window with an OK to
appear.
From the commands menu of the ICS panel, set the instrument mode to STANDBY.
The name tags of instrument functions will turn green.
From the commands menu of the ICS panel, set the instrument mode to ONLINE.
The name tags of instrument functions will turn purple as each function is initialised.
Quit the configuration tool and ICS panels.
Archive
the configuration file.