Nr. Document Section RIX Description Reviewed by Answer Answered by
1 User Requirements general I do not know whether I am supposed to review it. If so, I would have preferred D. Queloz and/or F. Bouchy to join the review board. I will in any case add some comments in the following. F. Pepe The science drivers are not really the subject of this design review. The mode was recommended by the IST in 2002 and the science drivers are only intended as short examples of possible applications L. Pasquini
2 User Requirements 3 Point 2 seems unrealistic. The precision obtained on m=17 stars is of the order of 30m/s in one hour! Asterosismology? Cluster giants -> Flames See 1. There is agreement that there are some strong cases, with emphasis on rel. faint objects with fairly low S/N.Practical results will depend on the scrambling efficiency which will only be known after practical tests. L. Pasquini
3 User Requirements 3 I see three possible domains of use of the HEF: 1) spectroscopy on faint stars. 2) Medium-precision RV on faint objects. 3) High-accuracy RV for all objects with better efficiency.
The asteroseismology and the search for extra-solar planets is part of 3). However, this might be in contrast with the chosen design, since the injection of the pupil into the fibers and the absence of a scrambler do not guarantee that the high accuracy is achieved. In view of that the scientific goals might be rephrased, and/or the design verified/re-considered.
F. Pepe, D. Queloz Agreed on the 3 domains. The design was chosen for ease of optical design and low cost. Scrambling is to be done by the fiber, efficiency to be seen. L. Pasquini/G. Rupprecht
4 User Requirements 4.1 Loss because of R reduction: 0% - 30 %, depending on the line-width of the stars. This might compensate the no-scrambler gain. F. Pepe correct, with respect to RV accuracy. The final balance will have to be investigated. G. Rupprecht
5 User Requirements 4.2 OHP is no using the scrambler as well, but you are right: Before using the scrambler the accuracy was about 15m/s. However, and in any case, a precision of 15 m/s is NOT acceptable. F. Pepe, D. Queloz We require the performance on bright objects to be much better than that. However, emphasis is on faint objects which have photon noise of the order of 10m/s and fibre scrambling should not deteriorate this. L. Pasquini
6 User Requirements 4.2 Based on the HARPS experience and Casse et al. We may expect a loss of scrambling by a factor of 5. Considering the guiding stability we would still expect a precision below 2m/s. However, this is only valid for star injection, and we do not have any experience how the scrambling behaves in case of pupil injection. F. Pepe, D. Queloz see 5. We will only know when we observe. L. Pasquini
7 User Requirements 4.3 I do not think that the pixel-sampling is limiting the precision in case of UVES and CES. F. Pepe agreed. L. Pasquini
8 User Requirements 4.3 I do not understand the sentence concerning "… the major source of error coming from the larger star area sampled". Could you please explain. F. Pepe agreed: not well formulated. A larger star area is more susceptible to contamination by the sim. Calibration (if used) L. Pasquini
9 User Requirements 4.3 From this section we deduce that the RV-precision goal for HEF is 10m/s. This is a lack of ambition for this mode. 10m/s can be achieved by "any other" instrument F. Pepe, D. Queloz agreed that we should be more ambitious. However, NOBODY (incl. UVES) has so far reached 10m/s at V=15 L. Pasquini
10 User Requirements 4.3 The resolution expected with the HEF is about 80'000. At this resolution the stellar lines are almost always resolved. Depending on the star the loss in RV precision can still be considerable, however. F. Pepe agreed; the loss in RV accuracy will come from loss in calibration accuracy L. Pasquini
11 User Requirements 5 There was an 6th option using the Kohler design for the light injection. In contrast to the present solution the star instead of the pupil would have been injected. For this solution the costs would have been higher (optics), but there would have been more space for the shutter mechanics F. Pepe There are 3 reasons why we chose pupil injection: 1) The pupil scrambles the position variations of the star. There are however second order variations of the angular flux distribution (far field) which affect pupil aberrations. One of the goals of this solution is to demonstrate the scrambling properties of the pupil injection. 2) There is only one air-glass surface instead of three in the case of the projection of the star on the fibre. Therefore the pupil solution is more efficient. 3) The cost of the single rod lens is around ten times less than the two doublets. G. Avila
12 User Requirements 5.5.1 I think we should also recall that changes have to be done to the shutter control, ICS/OS, templates, procedures. I agree that these do not have a direct impact on the instrument stability, but they should be underestimated. F. Pepe Agreed, they should not be underestimated and must be properly taken care of (by La Silla) before the mode is offered G. Rupprecht
13 User Requirements 5.4 This has to be regarded as a NEW PACKAGE and will require a significant work work (new keywords, templates, calibration and cal base, trigger, reduction, ETC, etc. F. Pepe agreed. See above. Input required from the Consortium. G. Rupprecht
14 User Requirements 5.5 I think that the Commissioning and test plan should be elaborated more in detail later (test of efficiency, scrambling/guiding, Dark current with closed shutter. F. Pepe agreed. This doc is a requirements and implementation plan, design docs and commissioning plan naturally come later. G. Rupprecht
15 Optics general Please discuss the alignment tolerances: a) position tolerance between hole, rod-lens, and fiber. b) Angular tolerance between rod lens and fiber axis. c) Angular tolerance between aixs of the fiber-lens system and the telescope axis. F. Pepe The tolerances have been defined by the FEROS fibres which are the same case as for HARPS: a) Eccentricity between the fibre end and the lens has to be less than 4μm. The centering between the lens and the hole in the mirror has to be better than 5μm for a vignetting of less than 3%. b) The angular tolerance between the lens and the fibre axis is quite large, it should be less than 1deg. This tolerance affects the field aberrations only. c) The angular tolerance between the hole-lens-fibre system with respect to the telescope pupil is adjusted by inspecting the projection of the fibre end on the M2 mirror. In the case of FEROS, the angular error has been estimated to be less than 0.5deg. This value gives a vignetting of the projection of the pupil on the fibre of thess than 2%. G. Avila
16 Optics general Please discuss the alignment procedure (without opening the vacuum vessel) F. Pepe We will adopt the splicing of the fibre in order to allow the back illumination for an optimal alignment. The efficiency has not been tested but the specifications of the splicer guarantee coupling efficiencies better than 97%. The procedure is laborious and time consuming but the principle is quite simple: 1) The fibre is mounted in the ferrule and polished. 2) The other end is also polished but without ferrule. 3) The end with the ferrule is mounted in a tilt stage and centered in front of a screen with concentric circles. This alignment is done by autocollimation on the ferrule. 4) The adjusting “three fingers” mechanical system made by Wolfgang is mounted on the ferrule and the lens is placed in front of the ferrule. 5) A laser is injected from the back and the projected pupil on the screen is centered with the help of the fine adjusting fingers. 6) Once the process is verified, the lens is dismounted and the glue is placed in front of the ferrule and the lens placed against. The alignment is done again and the setup is left several hours for curing. 7) A photodiode is placed in front of the out coming flux and the mirror diagram is placed in front of the lens. 8) The mirror – diaphragm is adjusted until get the maximum of flux in the detector. 9) Once the technique is tested, the mirror – diaphragm is dismounted to put the glue and repeat the adjustment. G. Avila
17 HEF for HARPS Please describe (if none) the open/close time for the shutter) Lizon ? (W. Eckert)
18 HEF for HARPS Can the SESO controller be used? Lizon ? (J. Alonso)
19 HEF for HARPS Can pulpo be used to control both shutters (two channels of puplo?) F. Pepe ? (J. Alonso)
20 HEF for HARPS I suppose the distance between the fibers is 16 mm. Is this correct? F. Pepe yes W. Eckert
21 HEF for HARPS You should foresee a time-out for the motor in case the shutter does not reach the final position. F. Pepe ? (J. Alonso)
22 HEF for HARPS Have following task been forseen: ICS/OS changes, Template changes. F. Pepe ? ?
23 HEF for HARPS The tolerence on the angular poisitioning might be very tight. Could you give some numbers and explain how to consider these tolerances correctly F. Pepe ? ?