Neutral density wheel calibration check

Version: 1.0 2007-07-31 glocurto

The fact:

On July 21st 2007 a problem was reported that triggered an intervention in the HARPS adapter. As a result the neutral density wheel was broken. The spare was installed and its calibration checked. This page documents the calibration test on the wheel.

The following table illustrates the correlation between encoder steps, filter density and number of counts in the exposure meter for few selected samplings. more measurements were taken, and the spread aroound the "theoretical" curve is within 20%. The only exception was the range of exposure times lower than 80s, where no attenuation is measured. This was due to a rounding problem now corrected. For the range of encoder steps from 11500 to 14400 there is no attenuation measured, while the filter density is estimated to be 2.15 by the system. This corresponds to a transparent sector of 91.25 degrees.

Encoder steps Filter density Exp. meter counts
0 0.0 245k
3900 1.0 23k
9600 2.0 2.1k