Optimum guiding for asteroseismology with HARPS

The optimum guiding strategy for asteroseismology may very well be different from the optimum strategy when looking for planets. In asteroseismology the biggest concern is to introduce as little periodic noise as possible in the relatively short exposures, while for planet hunting the noise should just be as low as possible with short periods being evened out by relatively long exposures.

We have conducted some experiments, where we have analysed intense series of short exposures on the same star and compared with a similar analysis of the various guiding and centering parameters. This document describes the results of these experiments.

Results

The following figure shows the power spectrum of two half nights of RV data on star1. Note the peak around 2.8mHz and the 1/f noise rising from around 3mHz and down.
The following plots show the power spectrum of the apparent position of the center of the fiber as calculated by the dynamic centering algorithm, i.e. it is the pixel the software uses as a reference to guide on. Left plot is X-coordinate, right is Y-coordinate. Note the peak around 2.8mHz and the high power at frequencies below 1mHz.

/-----/

The next figure is the power spectrum of half a night of RV data on star2. Note the peaks at 2.8mHz and 6.4mHz. Note also that the noise is essentially white down to 0.5mHz.
In this case the dynamic centering algorithm, which measures continuously the position of the fiber and corrects the guiding reference pixel, was turned off and small variations of the fiber positions were corrected by hand, only three times during the whole test (which lasted about 3 hours). The fact that the 2.8mHz peak is still present indicates that the dynamic centering is not responsible for it. The corresponding plot for star1 is essentially identical to the plot presented above.

/-----/

The following plot shows the power spectrum of the centroid position in X (left plot) and Y (right plot). The white noise level is about a factor of two higher than in the previous case (mostly due to the shorter series). Splitting of the series on star1 results in essentially the same noise levels.

/-----/

The last plot shows the guiding correction as actually applied during the series, i.e. the correction applied to the telescope to keep the star centered on the fiber. Note the very strong peak around 2.8mHz.

Discussion

From the presentation above we can conclude that the dynamic centering is not responsible for introducing any periodic noise in the RV measurements or the star's centroid position.

The most noticeable feature of the power spectra is a peak at 2.8mHz, corresponding to a period of about 6 minutes. This seems not to depend from the star as it was measured on two different targets. It should not depend either by the periodic nitrogen flow in the HARPS cryostat because the power at 2.8mHz was measured both in the RV data and in the star's centroid data as measured from the guide camera.

The only responsible for this peak could therefore be:

  • the guiding (althought it seems unlikely given that the guiding correction was applied every 5 seconds);
  • the telescope or the dome.

    After discussion with the La Silla mechanical engineers it became clear that the responsibles are either the alpha motor of the telescope, or the first gear (the one after the motor) of the alpha axis. Corrective intervention is under discussion.

    In the standard configuration, HARPS guiding delivers the lowest possible noise, although it is not entirely white, but has a strong 1/f or random-walk (drift) component. As this noise affects the typical range of interest for asteroseismology, the standard configuration is not suitable for seismology with HARPS.
    Based on the present experiments we recommend a much lower dynamic fiber centering correction frequency (15 minutes), which we have shown eliminates the 1/f component quite efficiently.