05-11-2006 earaya, glocurto ------------------------------ Current 10mA Fiber A THAR2 Fiber B THAR3 Lamp model 3UAX Th S/N counts frame 202 40k HARPS.2006-11-05T19:39:49.455.fits 205 84k HARPS.2006-11-05T19:43:07.725.fits 204 50k HARPS.2006-11-05T19:46:07.935.fits 201 35k HARPS.2006-11-05T19:49:10.334.fits 198 38k HARPS.2006-11-05T19:51:50.823.fits 199 85k HARPS.2006-11-05T19:54:20.000.fits 203 60k HARPS.2006-11-05T19:56:54.387.fits 207 54k HARPS.2006-11-05T19:59:22.036.fits Leave installed S/N B00199 Try to increase flux: -> 150k counts Change current of THAR3: I counts frame 8mA 114k HARPS.2006-11-05T20:23:21.719.fits 6mA 78k HARPS.2006-11-05T20:28:24.434.fits 9mA 140k HARPS.2006-11-05T20:32:04.725.fits 5mA 65k HARPS.2006-11-05T20:34:13.971.fits 7mA 98k HARPS.2006-11-05T20:36:09.227.fits 10mA 162k HARPS.2006-11-05T20:37:55.883.fits Current 10mA, move the calibration fibre carriage. Default position: 63736 (SPARE) Offset Counts File +1000 140k HARPS.2006-11-05T20:50:38.162.fits +2000 110k HARPS.2006-11-05T20:52:02.207.fits -1000 198k HARPS.2006-11-05T20:53:23.411.fits -2000 220k HARPS.2006-11-05T20:54:44.256.fits -3000 236k HARPS.2006-11-05T20:56:01.959.fits -4000 243k HARPS.2006-11-05T20:57:20.354.fits -5000 241k HARPS.2006-11-05T20:58:45.699.fits -6000 230k HARPS.2006-11-05T21:00:35.384.fits -7000 210k HARPS.2006-11-05T21:02:06.519.fits -8000 170k HARPS.2006-11-05T21:04:05.955.fits -9000 76k HARPS.2006-11-05T21:05:31.300.fits ============================================================ 06-11-2006 Testing other lamps (E. Araya, G. Lo Curto) Vacuum pump on Current: 10mA Instrument Producer S/N Counts File Coralie Oriel B47816 181k HARPS.2006-11-06T18:12:58.248.fits FEROS Juniper 528046 340k HARPS.2006-11-06T18:18:13.255.fits FEROS Juniper 528048 577k HARPS.2006-11-06T18:21:54.616.fits *** The FEROS lamps contains Ne as well (ThAr not available at the moment). Reinstall Heraeus lamp B00199 and align it: 140k. ============================================================ 07-11-2006 Testing Juniper (EMMI) lamp (A. Gilliotte, G. Lo Curto) S/N: 8490 Current: 10mA Lamp is too large to fit in the support, it is held by hand in front of the fibre. # of counts is varying between 1k and 340k. Files: HARPS.2006-11-07T19:16:57.634.fits possibly some contamination by a flashlight. HARPS.2006-11-07T19:18:16.489.fits