Critical issues during GROND operations This is just a short summary of the critical issues, a more extended version on other aspects of the GROND operations will follow within the next weeks. 1. NEVER SWITCH OFF GROND/FIERA WHILE THE CRYOSTAT IS COLD! The GROND vessel is cooled, via the CCC (closed-cycle cooler) to 80 K. The 4 CCDs are internally counter-heated to 165 K. The heaters and temperature sensors are controlled via FIERA/PULPO. IF FIERA is switched off, it takes about 10 min until the CCDs reach temperatures below 150 K, where the Silicon pixels will break off their boards - and all CCDs will be destroyed (we never tested this, but this is the statement of the manufacturers). 2. NEVER HIT THE TELESCOPE HARD WHEN THE TURBOPUMP IS RUNNING. The TMP (Turbo-Molecular-Pump) is presently running to completely evacuate the vessel. During later operations, it will run only about once per months, since the CCC acts as a nice cryo-pump. The running TMP (spins at 90.000 revolutions per minute!) only likes gentle movements (including telescope slews). However, it will be very likely destroyed by the "bangs" during telescope balancing. Therefore, the TMP should be switched off before telescope balancing, and one should wait about 15 min until the rotor stopped spinning (there is no break). 3. INFORM US IMMEDIATELY IF THEIR IS A POWER OUTAGE LONGER THAN 10 min At the moment, our emergency procedures are not yet running, so if there is a power problem (primarily with the 380 V or the UPS!) for longer than 10 min, please call me or any of our group at any time (also in the middle of the night!), so that we can run and counter-heat the detectors with our own UPS. In a later stage, our software should take care of this, so this is a temporary issue. during night during daytime Preferentially call: Gyula Szokoly, room 406, Tel. 4326 RITZ: 4552 Otherwise: Jochen, room 408, Tel. 4328 2p2 4108/9