Maintenance Intervention for FIERA J. Greiner, 25. May 2007 WARNING: NEVER SWITCH OFF GROND/FIERA WHILE THE CRYOSTAT IS COLD! Exception: Switch-off for 5 min is possible, if powered on afterwards, and heating of CCD detectors by PULPO is checked. If maintenance intervention is required to be done at the running instrument, the following procedure should be followed: 1) Switch-off GROND/FIERA Power 2) Disconnect power cable to GROND/FIERA Head - see IMG_9033.JPG 3) Switch-on GROND/FIERA Power 4) Perform maintenance operation. 5) Switch-off GROND/FIERA Power 6) Connect power cable to GROND/FIERA Head - see IMG_9034.JPG 7) Switch-on GROND/FIERA Power THE INTERVAL BETWEEN STEPS 1) - 3) or 5) - 7) SHOULD NOT EXCEED 5 MIN, OTHERWISE ALL 4 CCD-DETECTORS WILL BE DESTROYED! Example: In the following, the change of a Bias Board is described as example of a possible maintenance intervention: - Perform above steps 1) - 3) (IMG_9033.JPG) - Move up to centerpiece - most convenient place is standing between the margarida motors. - IMG_9028.JPG - Remove inner mounting leg of GROND/FIERA Head: Open 2 screws into centerpiece, and 4 screws connecting leg with GROND/FIERA Head box. - IMG_9029.JPG - Remove cover plate of GROND/FIERA Head: Open the 15 remaining little screws: - IMG_9030.JPG - Remove corresponding plugs; starting on the right side: open screws, and disconnect. - IMG_9031.JPG - Remove Bias board: open two screws at the side. Take off board. - IMG_9032.JPG - Insert new board, fix 2 screws at the side. - Connect the plugs. - Close the cover: insert 15 little screws. - Connect the mounting leg. Be careful not to squeeze the orange fiber! - IMG_9028.JPG - Perform steps 5) - 7) as above (within less than 5 min).