La Silla - Science Operation Department


FEROS daily checks

Version: 1.0 2002-11-06 Ivo Saviane

Every day

1
Check appearance of bad columns : from bias frame (aver/row on a median of 5 bias frames). It can signal too high pressure in the dewar.

2 2.5

morning and evening, check env monitor, at
/lasilla/instruments/feros/EnvMon/
-> T=17C? $\Delta T<0.2  C$ ?

3 2.2

are the two slices, of each fiber, evenly illuminated in FFs? If not, check the alignment of the image slicer

4

ThAr lamp should run at nominal current (e.g. $12  \mathrm{mA}$ , $10  \mathrm{mA}$ are used at Paranal). This increases lifetime.

5 2.7

Check that FF lamps are running at constant current

6 2.10

Check UT of BIAS PC and INS workstation

7 
Are slices aligned along CCD rows? If not, was CCD dismounted ``yesterday''?