When switching from FEROS to WFI observations it is now necessary to
ensure that:
The ADC is OUT
The SCSM is parked
The M3 is in WFI
Set the AAFT value to zero and click Apply in the TCS panel
Due to features of VLTSW, this has not yet been reliably configured as
an
instrument setup from the OS panel, therefore making these changes
should be
made from the ICS panel, as it has always been done for the M3 control.
To
do this depends on the situation the ICS is currently in:
ADC is IN
In this case the SCSM must already be in the PARK position, therefore
in the "FEROS ICS Control" panel:
Devices -> Deselect all devices
Select the ADCA device (click the radio button to the left of
the "adca" entry in the "ADC" tab.
From the adca's list button select "OUT"
Select the M3 device in the "Motors" tab.
From the M3's list button select "WFI"
Click on the Setup button at the bottom of the panel
Set the AAFT value to zero and click Apply in the TCS panel
Ready for WFI observations.
ADC is OUT
In this case the SCSM can be anywhere, therefore
in the "FEROS ICS Control" panel:
Devices -> Deselect all devices
Select the SCSM device (click the radio button to the left of
the
"scsm" entry in the "Motors" tab.
From the scsm's list button select "PARK"
Select the M3 device in the "Motors" tab.
From the M3's list button select "WFI"
Click on the Setup button at the bottom of the panel
Set the AAFT value to zero and click Apply in the TCS panel
Ready for WFI observations.
Switching from WFI to FEROS observations
When switching from WFI to FEROS observations it is only necessary to:
Set the AAFT value in the TCS to 15 and click Apply in the TCS panel