FEROS on the MPG/ESO-2.20m

Switching between instruments

Switching from FEROS to WFI observations

When switching from FEROS to WFI observations it is now necessary to ensure that: Due to features of VLTSW, this has not yet been reliably configured as an instrument setup from the OS panel, therefore making these changes should be made from the ICS panel, as it has always been done for the M3 control. To do this depends on the situation the ICS is currently in:

ADC is IN

In this case the SCSM must already be in the PARK position, therefore in the "FEROS ICS Control" panel:
  1. Devices -> Deselect all devices
  2. Select the ADCA device (click the radio button to the left of the "adca" entry in the "ADC" tab.
  3. From the adca's list button select "OUT"
  4. Select the M3 device in the "Motors" tab.
  5. From the M3's list button select "WFI"
  6. Click on the Setup button at the bottom of the panel
  7. Set the AAFT value to zero and click Apply in the TCS panel
Ready for WFI observations.

ADC is OUT

In this case the SCSM can be anywhere, therefore in the "FEROS ICS Control" panel:
  1. Devices -> Deselect all devices
  2. Select the SCSM device (click the radio button to the left of the "scsm" entry in the "Motors" tab.
  3. From the scsm's list button select "PARK"
  4. Select the M3 device in the "Motors" tab.
  5. From the M3's list button select "WFI"
  6. Click on the Setup button at the bottom of the panel
  7. Set the AAFT value to zero and click Apply in the TCS panel
Ready for WFI observations.

Switching from WFI to FEROS observations

When switching from WFI to FEROS observations it is only necessary to: Ready for FEROS observations.