Measuring the signal current (voltage) of a IK320 measuring system

 

  1. Measurement of layout checkpoints (plug P1, P2, position see supplement.)
  2. Connectors P1/P2 (P1 Signal axis Xl, P2 of signals axis X2):
  3. The signal amplitude at the pins 2 and 3 is determined by the current signal. The current/voltage conversion takes place with the factor 112mV/uAss.

Pin

signal assignation

1

U0-GND

2

U90

3

UOO

4

RI

 

TRIG.RI

6

GND

 

 

Example: UOO = 1.68 Vss (measured on pin 3 of P1 or P2)

I00 - U00/Factor = 1,68Vss/112mV/uA 15 uA

 

2. Measurement over the AD converter of the IK320

To access the AD converter see description chapter 5.2 data.

The analog values are stored under the Base address + $F8... $FE.

With slow speed now the MIN- and MAX- values of the signal amplitude are determined.

The current signal is calculated as follows: The amplitude Ass is the total amount of the MIN and of the MAX value. An AD transducer increment has 0,6lmV, from which the amplitude in Vss can be calculated. The amplification factor of the input stage altogether is 540mV/uA.

 

Example:

 

MAX AD-Converter: $67C8 arithmetic shift right 2 = $19F2
MIN AD— Converter: $984C arithmetic shift right 2 = $E613

 

Total amounts: $33df

Decimal: 13279

 

Uss = 0,61mV * 13279 = 8,1 Vss

Iss = 8,1 Vss / 540 mVuA = 15 uAss