Absolute Strip Encoders

Absolute strip encoders Heidenhain LIDA are installed in the alpha and delta axis.
The strip is mounted on a drum attached to the axis.

Specifications:

40 um marks
4096 interpolation on IK320 board
Alpha resolution: 0.001955539 arcsec/increment
Delta resolution: 0.002984261 arcsec/increment

Head replacement and adjustments:

  • To replace the head, protect first the head and the strip by introducing a plastic foil between the head and the strip. Unscrew the bolts holding the head in place and withdraw it.
  • Install the new head in its support, aligning it so its base is parallel to the strip. Adjust the distance using a plastic gauge of 0.7mm.
  • Connect the PWM8 encoder analyzer (Heidenhain) to a 12 to 30 volts power (special connector).
  • Connect the head to the INPUT connector of the PWM8 encoder analyzer by the special adapter cable that matches the head and the PWM8 box.
    Note: There are two plug in modules for the PWM8 box: one for 11 uA type head and one for voltage head. The 3.6m heads are "voltage" type.
  • Switch on the PWM8. Confirm that the option to supply +5 volts to the encoder head is on (Um:on)
  • Rotate the telescope axis and watch the reading amplitude of both signals (0° and 90°). The maximum amplitude is obtained by the optimal alignment of the head respect to the strip marks (fig. 1). To adjustment, release a bit the bolts that hold the head (A) and rotate it slightly searching the maximum signal amplitudes.
  • Match the amplitude of both signals by trimming the “Compensate” potentiometer in the head electronics and then adjust the amplitude to 1 Vpp with the “Strom” potentiometer.

    TVl, TV2 and Phase Adjustment:

  •  

    Tight the bolts and watch that the signals stay at the maximum.

     
     
    Fig. 1
    Fig 2.


     

    TVl, TV2 and Phase Adjustment:

      Reference mark adjustment (Ri signal)
    Use an oscilloscope with three channels and memory. Connect the 0° and 90° signals from PWM8 box to channels A and B respectively and add them. The third input is the reference itself. If you do not have a three-channel oscilloscope, you may add the 0° and 90° signals connecting them via two 200K ohm resistors to one channel. Connect the other scope channel to the reference mark signal (zero pulse). Set the trigger source to the reference mark signal, trigger mode to single and adjust the trigger level to +0.4 VDC, negative slope.

    Move the telescope and wait for the arrival of a zero pulse.

    The phase difference of the zero pulse (negative peak) respect to the sum of 0° + 90° signals should be less than 20° (fig. 4). To adjust the zero pulse phase, tilt the encoder head using the adjusting bolts in the head support. Don't get confused with the "spring tension" bolts!
    To adjust the zero pulse with, use the "Ri-Breite" potentiometer.

    Fig. 4
    Repeat all the previous adjustment