EFOSC Setup Guidelines


EFOSC Setup Guide Lines

Main Setup steps, procedures and tools

Always must be aligned all devices on the Grism and slit wheels

For Grism setup check the weights before install the device around the wheel

Look Grism weight
1.- Get the official setup request via Remedy Query(user/passwd required)
2.- Use efopanConfig panel to check and set the new configuration request. New added features are:
  • New entry field for comments or an extra optic element name. Useful for P2PP Special names.
    Example:
    For a filter P2PP name SPECIAL1 add in the comments field the ESO name like ESO#666
  • All the encoder and offset entries are no enable by default. Normaly this values should NOT be changed manually, but, for maintenance a select button was added to enable encoder entries.
  • New radio buttons by wheel position to enable both the optic "Element Name"(p2pp name list select option) and "Comments" fields. Once entries are done press button "SET VALUES" and confirm. Panel will be update at once. After changes never close the panel w/o doing the configuration backup under "file" menu options list and pressing the "RESTART ICS" button "
3.- Use Bob and select EFOSC_gen_tec_SetUp template collection to be used in the same sequence.
Templates Sequence:
4.- Open the efoopsSetupForm to display and log the instrument configuration and measuremnt results. Once setup is done write your name, cross-check the camera and collimator position. If there are differences just enter the right values and wait 15 sec to check same number in the display field next to the entry one, i.e. properly written to the DB. Finally DO PRINT and get one copy for the operator, one for the history folder and one for the user and of course signe all them.



3.1 -
EFOSC_gen_tec_FocwAlignment
    This procedure must be done always with the instrument in focus, i.e. after a Collimator true focus. (See Tilt v/s Temp). First step is to measure focw tilt and if it is bigger than 0.2 pixes a popup dialog box will ask you to
    "Increase the Collimator Position by adding xxx(mm) or DO the Collimator True Focus".
    The option to do the focus wedge alignment after confirm is still available just in the case the true focus template fail in doing its alignment.
    Again DO NOT ALIGN FOCW WITHOUT THE COLLIMATOR CORRECTION SUGGESTED EITHER BY THE FOCW ALIGNMENT OR THE TRUE FOCUS SEQUENCE TEMPLATES.
    Detector and Instrument setting are done by default, i.e. exposure time, window, speed, Holes_Mask, no filter and Foc_Wedge. Internal He lamp will be selected and controlled automically if the TCS is enable in the OS panel. If it's not, You must switch it ON manually before starting the template.
    The local offset grism wheel will be automatically written in the data base. Bob will hung-up ~2min while doing data base backup, so this is normal and is there to catch problems. Running in the background, wrong values could be detected very late during the observation.

    Back to Main
3.2 -
EFOSC_gen_tec_TrueFocus
  • Go to the Telescope, move a xterm near the instrument. Of course, connect it and do login to efosc@wefosc.ls.eso.org. If the midas was running in the RITZ console type efoopsStopMidas 11 and re-start Efosc from the user menu "INS Start".
  • As the telescope control will be STANDBY, TCS should be DISABLE in the EFOSC OS panel.
  • Switch ON He lamp by opening the Internal Lamps panel from the user menu options or try the following command:
    rsh $TCSENV -l tcs -e e3p6guiLamps -d $DISPLAY (In the near future should be nttGuiIntLamps)
  • Open the Instrument and write down the collimator position and current Efosc temperatures shown in the OS panel
  • In Bob select EFOSC_gen_tec_TrueFocus template and check default setting as follow:
    • DET => Exposure time 1 sec and window T and clock 6.
    • INS => Aperture wheel on Holes_Mask, other wheels Free and Halfwave plate out
    • SEQ => Focus sequence step number (default 11), Step (default 0.3), First Collimator position
    Suggested collimator range to be explore is from 0.5 to 3.5 millimeters, number of step >= 7 and maximun step size <= 0.4
    Check last measurements log here. Plots show the Focwedge dy v/s temperature, Collimator v/s temperature and Collimator v/s Focwedge dY. The slopes are:
    • Collimator ~ +0.1 (mm/C)
    • Foc_wedge ~ -0.133 (degrees/C) or 0.14 pix/C. So after 2(C) of changes or disalignment worse than 0.2pix it has to be re-aligned.
    The local offset grism wheel will be automatically written in the data base. Bob will hung-up ~1min while doing data base backup, so this is normal and it is there to catch problems. Running in the background, wrong values could be very late detected, i.e. during the observation.
    Back to Main
3.3 -
EFOSC_gen_tec_SlitsAlignment
  • The same procedure is usefull to align either the CCD or the slits.
  • Before mounting a MOS plate, use standard slit to align the corresponding aperture wheel position.
  • If TCS is anable and online, He internal lamp will be controled automatically, but, if it's not the case do it manually.
  • Select this template and set SLIT to "All" and press START
    All the mounted slits will be aligned automatically. The local offset stap wheel will be automatically written in the data base. Bob will hung-up ~2min while doing data base backup, so this is normal and is there to catch problems. Running in the background, wrong values will be detected during the observation.
    Back to Main
3.4 -
Grism Alignment
  • If TCS is anable and online, Quarz internal lamp will be controled automatically, but, if it's not the case do it manually.
  • Select this template and define the grism to be aligned and press START
    Also is posible to define "All", but, the exposure time by Grism table is not yet included.
  • The local offset grism wheel will be automatically written in the data base. Bob will hung-up ~2min while doing data base backup, so this is normal and is there to catch problems. Running in the background, wrong values will be detected during the observation.
    Back to Main
3.5 -
Polarization
  1. EFOSC_gen_tec_WollonGrism - (Image Polarimetry + Filters)
    This mode need the paralel wollaston mask installed on the slit wheel, but, to have it properly aligned a standard slit should be used to measure the best local offset for that position.
    Wollaston should go on grism wheel with vertical alignment better than 2 pixels. To align just follow the next steps:
    • If TCS is anable and online, He internal lamp will be controled automatically, but, if it's not the case do it manually.
    • Select template EFOSC_gen_tec_WollonGrism
    • Use default setting, i.e. HWP continuos rotation and IN, Holes_Mask(always with the polarizer plate in) and Wollaston.
    • Warning: midas will ask to mark the upper and lower image spot one after the other. DO NOT TRY BOTH in one iteration,  just read the midas console messages.
  2. EFOSC_gen_tec_WollonFilt - (Image Polarimetry w/o filter and SpectroPolarimetry)
    Image Polarimetry w/o Filters need the perpendicular wollaston mask installed on the slit, but, to have it properly aligned a standard slit should be used to measure the best local offset for that position.
    SpectroPolarimetry need Segmented slits on the aperture wheel which are defined putting the perpendicular mask together with any standard long slit in the same position. As there is a 1.5" segmented slit (with black tape) other segmented slit could be offered combining the perpendicular mask with any of the following standard slits: 0.5, 0.7, 1.0, 1.2, 2.0, 5.0, 10.0
    Wollaston should go on filter wheel with horizontal alignment better than 0.1 pixels. To align just follow the next steps:
    • If TCS is anable and online, He internal lamp will be controled automatically, but, if it's not the case do it manually.
    • Select template EFOSC_gen_tec_WollonFilt
    • Use default setting, i.e. HWP continuos rotation and IN, Holes_Mask(always with the polarizer plate in) and Wollaston.
    • Warning: midas will ask to mark the left and right image spot one after the other. DO NOT TRY BOTH in one iteration, just read the midas console messages.
  3. EFOSC_gen_tec_Polarimetry - (Polarization Test) use a pinhole + polarizador on the aperture wheel and the HWP and the wollaston to take pinhole images for different HWP angles in step of 22.5 deg. The idea is to find the angle where one of the spots signal is the maximun and in the other is the minimun. Adding or subtracting 90 deg. this situation will be reverted. To see the effect just use the RTD cuts trace line and two sine waves on midas graph display. Warning: There is a jump in the signal around 22.5 degress.
    Back to Main



3.6 -
Coronograph

Back to Main


Send comments to <ebarrios@eso.org>