[BOB Setup OBDs] First of all do not be in a hurry and easily take two hours after any light path exchange. Also is suggested to do it from the old CES control. Then go a head as follow:

  1. In the RITZ control desk quit both BOB and MIDAS PCO
  2. Take the TH atlas and go to the OLD CES control room
  3. Do login to the wces as user ces passwd ********. Open and xterm and type:
    bobMain & for BOB
    ceouifStatus & for Instrument status
    cesopsStartMidas for midas-pco
    cesopsSetupForm panel

    From BOB select the CES_gen_tec_CesSetup.obd template, which is in fact a collection of the needed scripts for a full instrument setup. Set the corresponding keywords, entrance slit width and the center wavelength in nanometers. Then selsct the brach to be executed:

  4. CES_gen_tec_SetWalen:Set Grating and Predisperser computed position as a wavelength function.
  5. CES_gen_tec_GraWaveCent: Fine Gratting re-centering on 2048.5 pixel (+/- 1).This is a semiautomatic procedure, Operator must define the wavelength lines position and type the numbers in the dialog box. If the template does either not converge or oscillate between two values, use the BOB panel Abort option. Then try this template CES_hcfa_cal_setwave.obd
  6. CES_gen_tec_PreWaveCent: Fine Predisperser re-centering on 2048.5 pixel +/- 5 This is an automatic procedure. Operator should check if the correction is oscillating between two smaller values like +/- 10 to abort the sequence and live it in the last position. This could happen in the red because fringes or ghost.
  7. CES_gen_tec_SlicerTilt: Slice Images have to be aligned with the CCD rows (cross-dispersion). This procedure will take a TH image and will compute the needed corrections to be manually applied. The Midas procedure will say when is OK.
  8. CES_gen_tec_CcdAlign: Spectrum images have to be aligned with the CCD columns. This procedure will take a FF image and will compute the needed corrections to be manually applied. Disalignment should be lower than 0.1 pixel.
  9. CES_gen_tec_SlicerTilt: If CCD was rotated, please DO the SLICER Alignment cross-check before checking Focus.
  10. CES_gen_tec_HartmannTest:Hartmann Test.
    Because the midas center/gauss command error to measure lines position is about 0.2 pixels, Hartmann test is only useful to define the starting focus umeter position for the through focus sequence.
  11. CES_gen_tec_TrueFocus:Through Focus.
    Define NSTEPS = 7 and STSIZE = 0.2 and set UPOSIT (focus umeter) = [ current_position + hartmann_test_suggestion - 1 ](mm). A family(one by detected line) of focus curve will be plotted. The best position will be obvious from it.
  12. CES_gen_tec_WidthVar: The slice image width variation in cross-dispersion is also measured. If a correction is needed and done, all the setup have to be checked e.i. repeat all the previous steps.( Once the above steps are done, width variation should be ok, if it's not the case, call opticien ...)
  13. CES_gen_tec_GraWaveCent: This is a second step, but it is suggested here after CCD alignment, Slicer and CCD table focus adjustment to verify the fine Grating positioning. Fine Gratting re-centering on 2048.5 pixel (+/- 1), is a semiautomatic procedure, Operator must define the wavelength lines position and type the numbers in the dialog box. If the template does either not converge or oscillate between two values, use the BOB panel Abort option. Then try this template CES_hcfa_cal_setwave.obd panel
  14. CES_gen_tec_ResPow: The last setup procedure is to measure the Resolving Power.
    This is a semiautomatic procedure, Operator must select two wavelength lines position and type the numbers in the dialog box. Midas procedure will compute the resolving power for the two entered wavelength and all the detected lines. Their wavelength will be computed from the linear dispersion function defined by the two entered points. Manual measurements can be done in the w3p6off using the last TH image.
    According to the user manual FWHM measurement for:
    • High-res should be in the range of 2.65 - 2.83 pix (540nm)
    • Medium-Res in the range of 4.15 - 4.62 pix (540nm)
    • Low-Resin the range of 6.23 - 7.79 pix (540nm)

    RP = [ line-wavelength(nm) / ( line-fwhm(pix) * linear-Dispersion(nm/pix) ) ]

  15. Use the cesopsSetupForm PRINT option, archive setup result in the control room folder and check Ces Setup History Plots