New M2 unit has three linear axis to adjust focus and collimation in X/Y direction. Their software range in mm are:
X/Y ~ 0.15 to 9.5 (mm) ; Center position 5.0/5.0 after initialization
Z ~ 0.00 to 32 (mm) ; Center position 13.4 after initialization
1 millimeters means 1000 encoders units
Focus and collimation can be corrected whenever be requested by the user (Particular corrections) as a function of the Serurier temperature and Zenith distance for Z and Telescope position model table for X/Y. Corrections can be done one by one or both. Open panel e3p6m2gui and see "Particular Correctionsns" buttons M21COR (both), FOC1COR (focus) and COL1COR (Collimation).
To do cycles corrections according to either frequency and/or position difference and/or serurier temperature variations criterias, check buttons START CYC(both) START FOC(focus) START COL(collimation). To stop them just press the corresponding STOP button.
Cycle corrections will be applied if telescope is in TRACKING, in any other state, it won't . Then for Telescope preset is not needed to stop it.
Focus formula Z(enc-units) = f1 + f2*SerurierTemp + f3*ZD can be tested just selecting the instrument and using the coefficients entry fields to be saved in the data base with the "Update DB" and computed with the "CALCFOC" buttons.
This is the Collimation model table entry for new x/y values by telescope position found after an Image Analisys Measurement and optimization.