*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 10 Oct 2003 13:46:16 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 10 Oct 2003 13:46:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#40 TAKEN ON 031010 IN 10 MODE ========================================================== BIAS ==== 5 biases at start test: average = 220.57 stddev = 7.70 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_10_40_031010 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 1558.3 1558.5 1246.7 0.10 0.10 1.25 2 8.00 3107.7 3108.4 2396.6 -0.08 -0.08 1.30 3 16.00 6208.1 6208.7 4802.4 -0.15 -0.15 1.29 4 24.00 9312.0 9310.6 7186.4 -0.14 -0.15 1.30 5 36.00 13969.5 13971.6 10592.8 -0.12 -0.10 1.32 6 48.00 18635.4 18634.8 14190.6 -0.06 -0.06 1.31 7 60.00 23305.0 23302.5 17688.9 -0.01 -0.02 1.32 8 80.00 31097.1 31096.8 23459.7 0.07 0.07 1.33 9 100.00 38905.3 38905.7 29302.0 0.16 0.16 1.33 10 120.00 46719.2 46718.6 34944.7 0.23 0.23 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2084 to 62478 e- Average count rate measured: 388 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.00 e-/ADU RON = 10.30 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.23/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay