*** MIDAS logfile (vers. 02FEBpl1.0) *** Th, 14 Aug 2003 22:51:15 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Th, 14 Aug 2003 22:51:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030814 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 384.3235 stddev = 1.9431 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030814 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 349.3310 349.0600 144.9920 1.0016 1.0009 1.20 2 6.799 2342.0161 2342.4460 960.7750 0.9961 0.9963 1.22 3 15.300 5276.0459 5277.0962 2158.6001 0.9980 0.9982 1.22 4 27.199 9387.9355 9386.9756 3890.5400 0.9992 0.9991 1.21 5 42.500 14679.175 14679.076 5822.6001 1.0000 1.0001 1.26 6 61.200 21153.675 21152.877 8282.2002 1.0008 1.0008 1.28 7 83.299 28807.377 28807.076 11170.349 1.0014 1.0014 1.29 8 108.799 37628.976 37627.976 14367.450 1.0015 1.0015 1.31 9 137.699 47609.375 47610.574 17560.099 1.0012 1.0013 1.36 10 170.000 58714.675 58714.574 20980.349 1.0002 1.0002 1.40 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.20 2 1.22 3 1.22 4 1.21 5 1.26 6 1.28 7 1.29 8 1.31 9 1.36 10 1.40 -------- --------- Dynamic range explored: from 968 to 162736 e- Average count rate measured: 345 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.77165E+00 +/- 2.83871E-03 e-/ADU RON = 5.38552E+00 +/- 5.51584E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** Th, 14 Aug 2003 22:51:39 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0016/ 0.9962 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay