*** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 11 Aug 2003 23:58:18 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 11 Aug 2003 23:58:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030811 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 383.4731 stddev = 2.1738 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030811 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 347.5330 347.3740 145.6765 1.0023 1.0018 1.19 2 6.799 2328.6670 2328.5869 953.1500 0.9962 0.9962 1.22 3 15.300 5246.7871 5247.0669 2153.5349 0.9982 0.9983 1.22 4 27.199 9330.8271 9331.9072 3811.2900 0.9989 0.9990 1.22 5 42.500 14594.226 14592.527 5838.6001 1.0000 0.9999 1.25 6 61.200 21028.326 21028.927 8407.5996 1.0007 1.0007 1.25 7 83.299 28637.427 28636.627 11141.500 1.0013 1.0013 1.29 8 108.799 37408.925 37409.527 14416.799 1.0014 1.0015 1.30 9 137.699 47330.628 47333.828 17470.800 1.0011 1.0012 1.35 10 170.000 58370.628 58372.527 21162.599 1.0001 1.0001 1.38 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.19 2 1.22 3 1.22 4 1.22 5 1.25 6 1.25 7 1.29 8 1.30 9 1.35 10 1.38 -------- --------- Dynamic range explored: from 953 to 160123 e- Average count rate measured: 343 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.74317E+00 +/- 2.96524E-03 e-/ADU RON = 5.96299E+00 +/- 6.44570E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 11 Aug 2003 23:58:42 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0020/ 0.9962 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay