*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 10 Aug 2003 10:37:31 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 10 Aug 2003 10:37:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030809 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 383.4604 stddev = 1.8374 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030809 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 345.6030 345.8860 142.4255 1.0010 1.0017 1.21 2 6.799 2318.9199 2319.4800 962.7300 0.9963 0.9964 1.20 3 15.300 5224.5801 5224.9600 2110.0801 0.9983 0.9982 1.24 4 27.199 9292.7002 9293.4004 3732.6799 0.9991 0.9990 1.24 5 42.500 14532.440 14533.040 5887.6001 1.0000 1.0000 1.23 6 61.200 20943.039 20941.439 8268.5498 1.0009 1.0007 1.27 7 83.299 28519.140 28520.640 11059.400 1.0014 1.0013 1.29 8 108.799 37253.539 37254.539 14039.099 1.0015 1.0014 1.33 9 137.699 47136.941 47138.039 17304.099 1.0013 1.0012 1.36 10 170.000 58133.339 58129.339 20963.949 1.0003 1.0001 1.39 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.21 2 1.20 3 1.24 4 1.24 5 1.23 6 1.27 7 1.29 8 1.33 9 1.36 10 1.39 -------- --------- Dynamic range explored: from 956 to 160765 e- Average count rate measured: 342 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.76554E+00 +/- 3.25336E-03 e-/ADU RON = 5.08142E+00 +/- 5.97772E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 10 Aug 2003 10:37:54 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0016/ 0.9963 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay