*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 10 Aug 2003 10:38:13 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 10 Aug 2003 10:38:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030809 IN ALOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 385.6193 stddev = 1.7725 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_ALOW_60_20030809 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 350.8500 350.5730 148.4785 0.9975 0.9968 1.18 2 6.799 2368.7710 2368.4109 973.9400 0.9989 0.9989 1.22 3 15.300 5332.0811 5332.2212 2205.8501 1.0001 1.0001 1.21 4 27.199 9481.8008 9481.4014 3859.0901 1.0006 1.0007 1.23 5 42.500 14818.481 14819.781 5991.4502 1.0010 1.0011 1.24 6 61.200 21341.380 21342.281 8263.3496 1.0012 1.0013 1.29 7 83.299 29045.980 29047.580 11292.650 1.0011 1.0013 1.29 8 108.799 37921.480 37922.882 14476.900 1.0007 1.0008 1.31 9 137.699 47962.281 47964.281 17786.500 1.0001 1.0002 1.35 10 170.000 59134.980 59136.179 21299.800 0.9988 0.9989 1.39 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.18 2 1.22 3 1.21 4 1.23 5 1.24 6 1.29 7 1.29 8 1.31 9 1.35 10 1.39 -------- --------- Dynamic range explored: from 966 to 162883 e- Average count rate measured: 348 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.75440E+00 +/- 3.11167E-03 e-/ADU RON = 4.88218E+00 +/- 5.51542E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 10 Aug 2003 10:38:35 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0012/ 0.9971 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay