*** MIDAS logfile (vers. 02FEBpl1.1) *** Th, 04 Sep 2003 17:02:09 Page 1 *** MIDAS logfile (vers. 02FEBpl1.1) *** Th, 04 Sep 2003 17:02:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030304 IN AABHIGH MODE ========================================================== BIAS ==== 5 biases at start test: average = 418.5517 stddev = 9.1240 There were 186 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_AABHIGH_60_20030304 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 el_adu2 -------- --------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 1741.9080 1740.3580 3342.0801 0.9985 0.9977 0.26 0.26 2 2.000 3447.4280 3445.5381 6386.2500 0.9979 0.9974 0.27 0.27 3 3.399 5848.7080 5849.9482 10999.400 1.0002 1.0005 0.27 0.27 4 6.099 10469.547 10471.047 19480.949 1.0004 1.0006 0.27 0.27 5 9.500 16300.948 16301.148 29980.449 1.0012 1.0013 0.27 0.27 6 13.699 23488.248 23491.347 42926.398 1.0011 1.0013 0.27 0.27 7 18.600 31874.347 31876.347 58500.500 1.0009 1.0011 0.27 0.27 8 24.300 41609.546 41612.546 75355.000 1.0004 1.0005 0.28 0.28 9 30.800 52701.949 52703.347 95233.500 0.9998 0.9999 0.28 0.28 10 35.000 59870.847 59865.847 107294.00 0.9996 0.9996 0.28 0.28 -------- --------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 970 to 33369 e- Average count rate measured: 1710 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 5.57370E-01 +/- 2.38832E-03 e-/ADU RON = 5.08544E+00 +/- 2.17910E-02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 2.00000E-02 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0012/ 0.9976 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay