*** MIDAS logfile (05SEPpl1.3) *** We, 27 Apr 2011 00:49:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.3) *** We, 27 Apr 2011 00:50:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#57 TAKEN ON 110426 IN norm MODE ========================================================== BIAS ==== 5 biases at start test: average = 56.01 stddev = 3.89 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_norm_57_110426 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 -------- --------- --------- --------- --------- --------- --------- 1 2.09 156.3 156.2 75.2 0.20 -0.09 2 3.99 295.8 296.0 141.9 0.06 -0.02 3 6.92 507.4 519.7 215.4 -0.61 0.61 4 13.99 1029.6 1029.7 410.3 -0.06 -0.13 5 27.99 2055.2 2055.7 809.8 -0.17 -0.21 6 41.99 3085.0 3084.7 1173.1 -0.07 -0.14 7 62.99 4623.3 4622.6 1729.5 -0.14 -0.22 8 83.99 6160.7 6160.7 2290.5 -0.19 -0.25 9 104.99 7711.7 7712.6 2806.7 -0.04 -0.09 10 139.99 10268.2 10264.5 3759.2 -0.18 -0.27 11 174.99 12857.5 12852.5 4622.3 0.00 -0.10 12 219.99 16191.7 16196.6 5792.9 0.18 0.15 13 259.99 19146.8 19128.1 6688.5 0.24 0.08 14 299.99 22115.2 22126.8 7667.6 0.34 0.34 15 309.99 22871.1 22868.6 7962.4 0.42 0.36 -------- --------- --------- --------- --------- --------- --------- Sequence el_adu1 -------- --------- 1 2.08 2 2.08 3 2.36 4 2.51 5 2.54 6 2.63 7 2.67 8 2.69 9 2.75 10 2.73 11 2.78 12 2.80 13 2.86 14 2.88 15 2.87 -------- --------- Dynamic range explored: from 451 to 65967 e- Average count rate measured: 73 ADU/sec. CONV. FACTOR AND RON ==================== *** MIDAS logfile (05SEPpl1.3) *** We, 27 Apr 2011 00:50:03 Page 3 C.Fact = 2.88 +/- 0.02 e-/ADU RON = 11.21 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.034 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.52/-0.44 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay