*** MIDAS logfile (vers. 95NOV) *** Th, 09 Dec 1999 15:11:36 Page 1 *** MIDAS logfile (vers. 95NOV) *** Th, 09 Dec 1999 15:13:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 991209 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1205.19 stddev = 6.81 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_991209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1262.7 1262.6 973.4 0.01 0.09 1.30 2 2.00 2515.2 2513.7 1899.6 0.00 -0.01 1.32 3 3.00 3764.9 3764.1 2769.3 -0.07 -0.07 1.36 4 6.00 7522.7 7519.7 5601.2 -0.03 -0.06 1.34 5 9.00 11272.2 11272.1 8277.0 -0.09 -0.09 1.36 6 12.00 15030.7 15030.6 10915.8 -0.06 -0.06 1.38 7 17.00 21302.4 21302.4 15525.9 0.00 -0.01 1.37 8 27.00 33841.3 33842.1 24414.7 0.04 0.04 1.39 9 37.00 46391.4 46384.6 33018.3 0.08 0.06 1.41 10 44.00 55188.5 55190.0 38789.7 0.12 0.12 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1792 to 78312 e- Average count rate measured: 1253 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.01 e-/ADU RON = 9.66 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12 /-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay