*** MIDAS logfile (vers. 95NOV) *** We, 01 Sep 1999 16:14:41 Page 1 *** MIDAS logfile (vers. 95NOV) *** We, 01 Sep 1999 16:16:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 990901 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1215.64 stddev = 6.54 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_990901 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1283.1 1284.3 981.3 -0.01 0.09 1.31 2 2.00 2559.3 2557.8 1948.3 0.02 -0.03 1.31 3 3.00 3832.3 3831.0 2847.3 -0.06 -0.09 1.35 4 6.00 7659.4 7656.9 5636.4 -0.03 -0.05 1.36 5 9.00 11483.7 11481.0 8384.2 -0.04 -0.06 1.37 6 12.00 15309.1 15308.9 11095.4 -0.04 -0.03 1.38 7 17.00 21687.8 21687.9 15748.1 -0.03 -0.02 1.38 8 27.00 34456.5 34453.5 24759.8 0.02 0.02 1.39 9 37.00 47239.5 47238.3 33910.2 0.07 0.08 1.39 10 44.00 56186.9 56187.5 39484.9 0.09 0.10 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1816 to 79503 e- Average count rate measured: 1276 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.41 +/- 0.01 e-/ADU RON = 9.25 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10 /-0.07 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay