*** MIDAS logfile (vers. 95NOV) *** Mo, 21 Dec 1998 14:35:26 Page 1 *** MIDAS logfile (vers. 95NOV) *** Mo, 21 Dec 1998 14:37:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 981221 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1102.01 stddev = 7.33 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_981221 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1345.7 1344.6 1050.5 0.06 0.07 1.28 2 2.00 2677.3 2677.6 2011.3 -0.07 -0.02 1.33 3 3.00 4012.0 4010.0 2989.1 -0.04 -0.06 1.34 4 6.00 8010.8 8009.5 5951.5 -0.07 -0.08 1.35 5 9.00 12012.5 12011.0 8841.4 -0.05 -0.06 1.36 6 12.00 16014.1 16015.4 11692.0 -0.05 -0.04 1.37 7 17.00 22689.9 22688.7 16435.1 -0.01 -0.02 1.38 8 27.00 36045.9 36047.0 25711.9 0.03 0.03 1.40 9 37.00 49424.1 49422.7 35331.0 0.09 0.08 1.40 10 44.00 58780.9 58779.2 41857.8 0.11 0.10 1.40 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1895 to 82804 e- Average count rate measured: 1334 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.41 +/- 0.00 e-/ADU RON = 10.33 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10 /-0.07 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay