*** MIDAS logfile (vers. 95NOV) *** Mo, 07 Dec 1998 15:06:08 Page 1 *** MIDAS logfile (vers. 95NOV) *** Mo, 07 Dec 1998 15:07:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 981207 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1118.78 stddev = 6.72 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_981207 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1347.9 1347.6 1038.7 0.00 -0.02 1.30 2 2.00 2686.5 2687.8 2029.5 0.01 0.05 1.32 3 3.00 4025.6 4024.4 3002.4 0.02 -0.01 1.34 4 6.00 8033.8 8034.2 5903.1 -0.08 -0.08 1.36 5 9.00 12048.8 12050.3 8865.1 -0.06 -0.04 1.36 6 12.00 16065.1 16063.6 11709.0 -0.04 -0.05 1.37 7 17.00 22757.1 22757.9 16590.1 -0.03 -0.02 1.37 8 27.00 36152.7 36151.9 26070.0 0.01 0.01 1.39 9 37.00 49560.7 49562.3 35548.1 0.06 0.06 1.39 10 44.00 58955.2 58956.4 41968.0 0.09 0.09 1.40 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1893 to 82797 e- Average count rate measured: 1338 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.40 +/- 0.00 e-/ADU RON = 9.44 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09 /-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay