*** MIDAS logfile (vers. 95NOV) *** Tu, 01 Dec 1998 19:08:24 Page 1 *** MIDAS logfile (vers. 95NOV) *** Tu, 01 Dec 1998 19:09:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 981201 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1119.13 stddev = 6.71 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_981201 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1379.6 1379.1 1060.9 0.02 0.07 1.30 2 2.00 2746.7 2747.0 2068.7 -0.03 0.01 1.33 3 3.00 4113.5 4113.8 3058.6 -0.06 -0.04 1.34 4 6.00 8215.2 8214.5 6114.4 -0.07 -0.08 1.34 5 9.00 12319.4 12320.6 9012.7 -0.05 -0.05 1.37 6 12.00 16426.3 16424.6 11955.4 -0.03 -0.05 1.37 7 17.00 23268.9 23271.0 16942.1 -0.02 -0.02 1.37 8 27.00 36967.7 36967.7 26507.5 0.03 0.02 1.39 9 37.00 50682.6 50679.5 36517.2 0.09 0.06 1.39 10 44.00 60288.3 60285.2 41876.9 0.12 0.09 1.44 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1964 to 85822 e- Average count rate measured: 1368 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.01 e-/ADU RON = 9.55 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11 /-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay