*** MIDAS logfile (vers. 95NOV) *** We, 22 Mar 2000 15:36:12 Page 1 *** MIDAS logfile (vers. 95NOV) *** We, 22 Mar 2000 15:37:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 000322 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1213.05 stddev = 6.57 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_000322 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1269.8 1271.5 983.2 0.00 0.04 1.29 2 2.00 2532.1 2533.0 1899.1 0.00 0.00 1.33 3 3.00 3792.6 3793.0 2813.4 -0.04 -0.06 1.35 4 6.00 7578.5 7576.8 5547.1 -0.03 -0.06 1.37 5 9.00 11363.2 11363.1 8270.8 -0.04 -0.04 1.37 6 12.00 15149.3 15148.6 11058.6 -0.03 -0.04 1.37 7 17.00 21458.8 21460.5 15547.2 -0.03 -0.02 1.38 8 27.00 34091.1 34090.1 24185.0 0.01 0.01 1.41 9 37.00 46737.2 46741.3 33549.6 0.06 0.07 1.39 10 44.00 55599.2 55598.7 39312.2 0.10 0.10 1.41 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1794 to 78517 e- Average count rate measured: 1262 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.41 +/- 0.01 e-/ADU RON = 9.28 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10 /-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay