*** MIDAS logfile (vers. 95NOV) *** We, 09 Feb 2000 14:04:24 Page 1 *** MIDAS logfile (vers. 95NOV) *** We, 09 Feb 2000 14:05:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 000209 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1199.77 stddev = 6.92 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_000209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1249.4 1248.7 969.7 0.02 0.06 1.29 2 2.00 2489.1 2489.3 1882.7 -0.07 -0.02 1.32 3 3.00 3733.0 3732.1 2793.8 0.01 0.01 1.34 4 6.00 7459.4 7458.3 5473.6 0.02 0.02 1.36 5 9.00 11174.3 11172.8 8090.6 -0.07 -0.09 1.38 6 12.00 14898.7 14895.7 10910.1 -0.06 -0.08 1.37 7 17.00 21108.5 21107.7 15221.4 -0.04 -0.04 1.39 8 27.00 33549.7 33547.0 24324.2 0.05 0.04 1.38 9 37.00 45968.9 45968.2 32852.5 0.04 0.03 1.40 10 44.00 54692.8 54689.1 38647.5 0.09 0.08 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1762 to 77172 e- Average count rate measured: 1242 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.41 +/- 0.01 e-/ADU RON = 9.77 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09 /-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay