*** MIDAS logfile (vers. 95NOV) *** We, 19 Jan 2000 20:28:23 Page 1 *** MIDAS logfile (vers. 95NOV) *** We, 19 Jan 2000 20:29:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#01 TAKEN ON 000117 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 1205.02 stddev = 6.92 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_01_000117 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1285.5 1286.8 986.0 0.07 0.08 1.30 2 2.00 2560.1 2560.5 2401.0 -0.01 -0.03 1.07 3 3.00 3834.1 3834.1 2877.6 -0.05 -0.07 1.33 4 6.00 7658.7 7658.6 5610.1 -0.06 -0.07 1.37 5 9.00 11483.5 11483.8 8345.2 -0.06 -0.06 1.38 6 12.00 15310.7 15309.7 11155.4 -0.04 -0.05 1.37 7 17.00 21689.8 21690.8 15889.8 -0.03 -0.02 1.37 8 27.00 34459.7 34459.6 24722.9 0.02 0.03 1.39 9 37.00 47241.9 47241.2 33884.1 0.07 0.07 1.39 10 44.00 56196.3 56196.4 39729.3 0.10 0.11 1.41 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1820 to 79537 e- Average count rate measured: 1276 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.01 e-/ADU RON = 9.79 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.007 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10 /-0.07 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay