- 1998 to mid-2005:
Gain (top panel) and read-out-noise (bottom panel) evolution for the SofI detector.
The solid horizontal line at 2 ADU is drown to guide the eye demonstrating a decrease of the RoN.
The images are clickable, linked to eps files.
Same as above, except the solid lines are fits for the data after mid-2006 (marked with a vertical
dashed line), and the dotted lines show the +/-1-sigma deviations (only some are shown on the bottom
panel for clarity). There is a tendency toward increasing the readout noise.

- 1998 to mid-May 2011:
Same as above, except the solid lines are fits for the data after mid-2006 (marked with a vertical
dashed line), and the dotted lines show the +/-1-sigma deviations (only some are shown on the bottom
panel for clarity). The gain is stable but the tendency for increasing readout noise that was seen a
year ago, is confirmed.

- 1998 to early-May 2012:
Same as above, except the solid lines are fits for the data after mid-2006 (marked with a vertical
dashed line), and the dotted lines show the +/-1-sigma deviations (only some are shown on the bottom
panel for clarity). The gain is stable but the tendency for increasing readout noise that was seen a
year ago, is confirmed.

- 1998 to early-Sept 2012:
Same as above, except the solid lines are fits for the data after mid-2006 (marked with a vertical
dashed line), and the dotted lines show the +/-1-sigma deviations (only some are shown on the bottom
panel for clarity). The gain is stable. The tendency for increasing readout noise that was seen a
year ago may be flattening.

Nota Bene: The files bellow contain occasional spurious measurement. Ignore the values that differ too much from the average ones.