Tests with new ThArNe lamP
Version: 1.0 2003-03-05 Ivo Saviane
Introduction
The FEROS calibration arcs are obtained by the superposition of a Thorium
lamp filled with Argon (ThAr), plus a Neon lamp. On February 22, 2003, I
replaced the ThAr lamp with a Thorium lamp filled with both Ne and Ar (ThArNe),
in the proportion Ne:Ar=9:1. I switched off the Ne lamp, and took spectra
of the ThArNe lamp. Wavelength calibrated line lists (obtained with the
@@ focus procedure) can be downloaded as well.
Raw and extracted data
You can download the raw and extracted+merged spectra by clicking on one
of the links below. All files are FITS gzipped. Raw 2d spectra are from 7.4
to 9.4 MB. Extracted+merged orders are ca. 700 KB.
10 sec raw extracted: object / sky line lists: object / sky
30 sec raw extracted: object / sky line lists: object / sky
60 sec raw extracted: object / sky line lists: object / sky
120 sec raw extracted: object / sky line lists: object / sky
A few tests
Below I show the 60 sec echelle spectrum of the new lamp, compared to that
of the ThAr lamp, and that of the ThAr + Ne lamps (same exposure time, same
low and high cuts). Just looking at the new lamp one clearly sees that it
is brighter and has a pinkish color, while the old lamp is fainter and has
a blue/violet color. The spectra below indeed show that the new lamp provides
a higher flux and it has many more red lines.
ThArNe 60 sec (new lamp)
|
ThAr 60 sec (just one old lamp)
|
ThAr and Ne 60 sec (both old lamps)
|
|
|
|
These qualitative remarks are confirmed by the plots below
Plot of row 2050 of new lamp spectrum
|
Plot of row 2050 of old lamps spectrum
|
|
|
Plot of extracted order 35, corresponding 7657-7946
AA
|
Same order for the old lamps
|
|
|
And finally, you can compare the results of the
@@ focus procedure
here:
Result of @@ focus, using new lamp
|
Results of @@ focus, using old lamp
|
|
|