New CCD Clock Patterns
New CCD clock patterns installed on 2004-Apr-23. These new clock
patterns correct a problem that caused the images from the FIERA CCD
controller to have 48 pre-overscan pixels and 52 post-overscan pixels,
instead of the normal 50 and 50. This is not a major problem, really
only affecting only reduction software, but easy to account for (as is
done automatically in the official FEROS-DRS
package). This problem affects ALL images acquired between the dates of
2003-October-01 and 2004-Apr-23.
The following images show that the new patterns indeed result in the
correct image format.

Figure 1: plot of row 2000, columns -50 -- 2. The images are
CCDTest flatfields. The blue line is an image from immediately before
the installation of the new clock patterns, the red line an image from
immediately after. Clearly we see that the new clock patterns give
SCIENCE data starting in column 0 instead of column -2.

Figure 2: plot of row 2000, columns 2045 - 2098. The images are
CCDTest flatfields.
The blue line is an image from immediately before the installation of
the new clock patterns, the red line an image from immediately after.
Clearly we see that the new clock patterns give SCIENCE data untill
in column 2048 with the overscan pixels starting in column 2049 instead
of column 2047.

Figure 3: plot of row 2000, columns 1250 - 1350. The images are
CCDTest flatfields.
The blue line is an image from immediately before the installation of
the new clock patterns, the red line an image from immediately after.
This image shows the bad column which is nominally in the 1300th column
of the SCIENCE data.

Figure 4: plot of row 2000, columns 1250 - 1350 (here the x-axis
is in units of mm as set by the FEROS-DRS prered.prg script). The
images are CCDTest
flatfields.
The blue line is an image from immediately before the installation of
the new clock patterns, the red line an image from immediately after.
This image shows that the DRS software properly 'corrects' the bad
column which is nominally in the 1300th column
of the SCIENCE data. The correction of this (and other) bad column is
made by averaging neighbooring pixels.
The results of the CCDTests carried out immediately before and after
the installation can also be compared. CCD tests were done with the RED
LEDs in both
225kHz,1,low
and
625kHz,1,med
readout modes. Comapre the two sets of results obtained on 2004-04-23.
Image Format: Difference between BIAS and FIERA images
Note that although FIERA images now have the correct format (50 pre and
50 post overscan columns and 2048 SCINECE columns) they still have a
different pixel cordinate system from BIAS images. The 'lower left
corner' of BIAS pixels has cordinates 1,1 and SCIENCE data is in
columns 51 to 2098, where as for FIERA images, the 'lower left corner'
has cordinates -48,1 and the SCIENCE data is in columns 1 to 2048.
Again the official
FEROS-DRS
package automatically correctly deals with these differences between
BIAS and FIERA images.
What are BIAS and FIERA
The Brorefeld Image Acquisition System or BIAS is the original CCD
controller used with FEROS until 2003-Oct. In 2003-Oct, the CCD
controller was changed to an ESO standard FIERA controller.