Flexure in the FEROS/WFI Adapter between FibreHead and Technical CCD
A study of the possibility of flexure of the FEROS/WFI Adapter (FWA)
between the FibreHead (FH) and Technical CCD (TCCD) has been made. The
result is that there no significant
flexure.
Description of tests
On 2003-07-17 the telescope was pointed first at the Zenith, then to
extreme values of Hour Angle (HA) and Declination (see table 1). At
each of the five pointings an image was captured from the TCCD (simply
using the
Save As... menu item from the
File menu in
the
TCCD RTD window) with the FEROS Calibration Unit (FCU)
Wavelength Calibration lamps turned on. The resulting images where then
analysed using the
FFH-TCCD-FlexureTest-plots.prg
MIDAS script. To summarise this procedure, 25×5 pixel regions
centered on the
expected positions (i.e. those determined at
the zenith) of the Object and Sky Fibres are summed over columns and
rows separately to produce a 1-dimensional summed flux profile, which
is "approximately guassian with a hole in the middle". The minimum of
this profile at the center of the Gaussian is interpretted as the
center of the fibre.
Results
Below are plotted row and column profiles of the flux summed over the
five pixels centered on the expected positions of the fibres (left) and
also the profile across the single row/column of the expected position
of each fibre (right) for both fibres.
As can be seen in the plots of the profiles of the summed flux for each
of the five pointings (except in the case of the column profile of the
Object Fibre at HA=00:00, Dec=+21:00) the center of the fibre
moves by somewhat less than 1 pixel , i.e. less than 0.51 arcsec (for
the image scale of the TCCD at the time these tests were performed). The
solid vertical line is drawn at the position of the expected centre, the
dotted lines represent the limits of the aperture (1.8 arcsec apertures
and image scale 0.51 arcsec/pixel implies apertures are 3.53
pixels).
Data
The image data files can be downloaded here:
FFH-TCCD-FlexureTest-plots.tgz.