Performance
Quality Control
|
norm
|
slow
|
fast
|
norm
2x2 binning
|
slow
2x2 binning
|
Daily Health
Check |
225kHz,1,low,1x1 |
N/A |
N/A |
225kHz,1,low,2x2 |
N/A |
Standard
Calibrations |
225kHz,1,low,1x1 |
60kHz,1,high,1x1 |
625kHz,1,med,1x1 |
225kHz,1,low,2x2 |
60kHz,1,high,2x2 |
Instrument
Efficiency |
225kHz,1,low,1x1 |
60kHz,1,high,1x1 |
625kHz,1,med,1x1 |
225kHz,1,low,2x2 |
60kHz,1,high,2x2 |
CCDTests are here.
Daily Health Check
Since 2005-Nov-19, as part of the Calibration
Plan,
a Daily Health Check OB consisting of:
- In 225kHz,1,low,1x1 readout mode
- 2 BIAS exposures
- 2 FLAT exposures
- 1 ThAr+Ne exposure
- In 225kHz,1,low,2x2 readout mode
- 2 BIAS exposures
- 2 FLAT exposures
- 1 ThAr+Ne exposure
is executed each afternoon. This OB allows the verification of the
health of
the instrument and lamps as well as permitting an initialisation of the
FEROS DRS for both 1x1 and 2x2 readout modes.
Standard Calibration Sets
As part of the Calibration Plan,
sets of Standard Calibration consisting of:
- 5 BIAS exposures
- 10 FLAT exposures
- 6 ThArNe exposure (3 different exposure times)
- 6 ThAr+Ne exposure (3 different exposure times)
are acquired each morning in each readout mode used during the night.
Instrument Efficiency
As part of the Calibration Plan,
the instrument efficiency, or more
precisely the instrument throughput is measured for every night with
FEROS
SCIENCE observations via the observation of Spectrophotometric Standard
stars.
ADC efficiency
The black curve shows the ADC efficiency. The transmission shows a decrease below 400 nm.
Therefore, a new ADC with a better trasnmission in the blue part of the spectrum (red curve) has been orderer and is foreseen to replace the current ADC from
period 78 on.