Imaging Science Modes

Contents


The 4 science imaging modes are described below. The observer is also referred to the pages on Acquisition images and Miscellaneous information.

Simple Imaging

In this mode one can obtain one or more images of a field through broad and/or narrow band filters.

The following acquisition templates may be used in this mode:

A sample jP2PP observing block

The above example shows an OB comprising a sequence of:

The FITS file is called EFOSC_Image.#.fits

Parameter Values One does not have to (in fact, should not) change the value of any parameter other than :

Jittered Imaging

This is very similar to Simple Imaging  except for the folowing difference :

A sample jP2PP observing block (the OB has been erroneously named Coronographic imaging)

The above example shows an OB comprising a sequence of

The FITS file is called  EFOSC_ImaJit.#.fits

Parameter Values   In addition to the parameters listed under Simple Imaging the following may need to be changed :

Coronographic Imaging

The Coronographic mode is very similar to the Simple Imaging Mode  except for the following differences :

A sample jP2PP observing block The FITS file is called  EFOSC_ImaCor.#.fits

Parameter Values  see under simple imaging

Polarimetric Imaging

Note that since EFOSC2 was moved to the Nasmyth focus of the NTT there is a significant instrumental polarisation that varies with telescope pointing. We are working to characterise this effect, but users should be aware that it may limit the accuracy to which polarisation can be measured. When preparing proposals for polarisation, extra time for observation of a large number of calibration stars should be included.

The acquisition template to be used is Img_Acq_Pol.

The imaging is similar to Simple Imaging except for the following additional features :

There is also a quarter wave plate for measuring circular polarisation (as opposed to linear polarisation with the HWP). Only one wave plate can be mounted at a time, as they occupy the same position in the instrument. Observing with the QWP is the same as with the HWP, but the QW versions of the templates are used.

Refer to Polarisation optics for more details.

IMPORTANT When the half wave plate (HWP) is set to continuous rotation in a template the counter ends up with 360xN degrees. When a signal is sent to set the HWP to some small fixed value (say, 45 degrees)  it tries to undo all the N rotations and in doing so triggers a time-out error. So we have included a command at the end of every Polarimetry template to initialise the HWP if it was set into continuous rotation mode - this just takes a few seconds. However when such a template is aborted, for whatever reason, this initialisation does not take place. The only solution is to cold-start the EFOSC2 instrument which takes about 1-2 minutes. So avoid aborting a Polarimetry template - use the STOP option instead.

A sample jP2PP observing block

The above example shows an OB comprising a sequence of: