EFOSC2 

ESO Faint Object Spectrograph and Camera
 
 
  
Dome/Sky/Internal Flats -- Spectroscopy

Spectroscopic Dome/Sky Flats
Spectroscopic Internal Flats

You find here a discussion of the  different types of spectroscopic flats.

Spectroscopic Dome/Sky Flats

Spectroscopic Dome and Sky flats are defined using the same templates and observed in an identical manner; i.e. the telescope is not moved from one spectroscopic sky flat to another. The only difference is the purpose they serve :  sky flats are used to determine the slit transfer function while dome flats are used to determine the pixel-to-pixel variation of the CCD gain.

One only needs to define the required number of counts in the template. The procedure takes a 1 second exposure and calculates the peak counts (after subtracting the bias and excluding pathological pixels). It then calculates the exposure necessary to achieve the required counts. If the required time is less than 1 second or more than 300 seconds the procedure returns an error and aborts the sequence.

Dome flats must be completed during the afternoon calibration session between 4.30 PM (local time) and sunset. The illumination source consists of an array of lamps of staggered brightness values. Different grism/binning/slit combinations require different settings of the lamps to achieve the required flat field level within a sensible amount of time (5 - 75 seconds). One can string together all the templates, with different grisms, slits and binning modes, in the same OB. However it would be more practical if one  arranges the templates in an ascending (or descending) sequence of the required lamp brightness - a good indication of which being the grism dispersion.

jP2PP Panel with Typical Parameters

The above example shows an OB comprising a sequence of:

The FITS image output by the template is named  EFOSC_FlatSpec.#.fits

As an example the following image is given, which shows a spectroscopic Dome Flat of 40,000 ADU peak counts with Grism #5, Slit 1.0", normal readout and 2x2 binning. Click on the image to get a larger version:

Parameter Values  One does not have to (in fact, should not) change the value of any parameter other than :

Spectroscopic Internal Flats

Internal Flats are only required when one wants to take flat frames while pointing to the target during the night. We have seen on a few occasions that grisms with heavy fringing in the red are not adequately defringed by the dome flats. This may be due to a combination of strong and narow fringing and CCD distortion with pointing. This effect is most obvious in Grism #12.

jP2PP Panel with Typical Parameters

The above example shows an OB comprising a sequence of

The FITS image output by the template is named  EFOSC_FlatIntSpec.#.fits

Parameter Values   One does not have to (in fact, should not) change the value of any parameter other than :

back to top