[ESO Web]
European Southern Observatory
New Technology Telescope
[NTT Web]

General :
Observations :
Operations :

Recovery Procedures

Procedure:
SOFI Grism Wheel - Initialization failure.
Author:
Ueli Weilenmann. 
Release:
1.0/December 23, 1999


If an init on the grism function fails, do the following:
------------------------------------

The points 1-6 are listed with increasing severity of the problem, so always start with 1 and only if this fails go to the next point.

PLEASE NOTE DOWN ALL THE STEPS AND THEIR RESULTS, in order to backtrack the bug later.

1. Check if the wheel turns.

2. Reboot the LCU 'lsofi' and restart the session. 3. Switch off and then reboot the LCU 'lsofi' and restart the session. 4. Move the wheel into the open position. 5. Move the wheel with the LCU handset into the open position and disable the grism function AGAIN: Please note carefully down what you do step by step!
 
 

 

[Top of Page]

Send comments to: NTT Webmasters <nttweb@www.ls.eso.org>
Last Update: MMM DD, YYYY (WHO)
[ESO][ESO][ESO][Search][Help][News]