After Image analysis, you find you have to adjust telescope focus by 0.2mm to have the instrument in focus.
The probable reason is that adapter was rebooted since Image Analysis software was started:
Therefore apply the following procedure in this order:
- on AO focus offset panel
- select the instrunent
- INITAO
- REFIA
- Check Focus again.
It is critical that the trombone is set at the position corresponding to the correct instrument BEFORE doing the REFIA. If not, REFIA stores in memory the position of the spots corresponding to a wrong focus and the following image analysis will try to go back to that last wrong focus.